inforMED
MalfunctionFSO

OPMI LUMERA 300

Received Dec 22, 2023 · Event occurred Dec 1, 2023

Report 3008564898-2023-00005 · MDR key 18384204

Device

Generic name

Microscope, Surgical, General & Plastic Surgery

Model number

300

Catalog number

000000-2098-818

Product problems

  • Component Missing
  • Component Missing

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ROOT CAUSE: THE MISSING SCREW WAS NOT ASSEMBLED DURING THE ASSEMBLY PROCESS OF THE SUSPENSION ARM.

Description of Event or Problem

DURING REPAIR OF THE OPMI LUMERA 300 DEVICE, A ZEISS FIELD SERVICE ENGINEER (FSE) FOUND THAT A SCREW WAS MISSING FROM THE SUSPENSION ARM AND THE YOKE CONNECTOR WAS STUCK ON THE SHAFT OF AXIS 4. THE FSE REPAIRED THE DEVICE BY REPLACING THE YOKE CONNECTOR AND PUT IN THE REQUIRED SCREW IN THE SUSPENSION ARM. THERE WAS NO INJURY TO A PATIENT OR THIRD PARTY.