BUBBLE SENSOR DETECTOR
Received Dec 21, 2023 · Event occurred Nov 29, 2023
Report 9611109-2023-00648 · MDR key 18370656
Device
Generic name
Monitor And/or Control, Level Sensing, Cardiopulmonary Bypass
Manufacturer
Livanova DeutschlandModel number
23-07-50Product problems
- Over-Sensing
- Failure to Sense
- Over-Sensing
- Failure to Sense
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A BUBBLE SENSOR DETECTOR DID NOT DETECT AIR BUBBLES. THE ISSUE OCCURRED DURING PROCEDURE. THERE WAS NO PATIENT INJURY.
Additional Manufacturer Narrative
A.1.-A.5. THERE WAS NO PATIENT INVOLVEMENT. H10: LIVANOVA DEUTSCHLAND MANUFACTURES THE BUBBLE SENSOR DETECTOR. THE INCIDENT OCCURRED IN LINCOLN, NEBRASKA. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
Additional Manufacturer Narrative
H10: THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT BUBBLE SENSOR BLADDERS WORN OUT. THEREFORE, IN ORDER TO SOLVE THE ISSUE, THE CUSTOMER ITSELF REPLACED THE ENTIRE BUBBLE SENSOR DETECTOR. BASED ON THE INFORMATION RECEIVED, THE ISSUE IS RELATED TO DEFLATED BLADDERS OF BUBBLE SENSOR WHICH CAN RESULT IN OVER-SENSING. IN CASE OF BUBBLE SENSOR TRIGGERING FALSE BUBBLE ALARMS, THE PUMP IS STOPPED BY THE BUBBLE ALARM EVEN IF NO AIR IS PRESENT IN THE MONITORED LINE. IN SUCH SITUATIONS, THE ALARM CAN ALWAYS BE CLEARED/OVERRIDDEN BY USER, ONCE THE LINE IS VERIFIED TO BE FREE OF AIR, AND THE PERFUSION CAN BE EASILY RESTARTED. THUS, IT IS UNLIKELY THAT A FALSE BUBBLE ALARM WILL RESULT IN PATIENT SERIOUS INJURY OR DEATH. THEREFORE, THE EVENT HAS BEEN RE-ASSESSED AS NOT REPORTABLE.
Description of Event or Problem
SEE INITIAL REPORT.
Remedial action
- Other