MalfunctionILY
104 LED CLUSTER PROBE WAVELENGTH = 660NM & 850NM ISSUE 1
Received Dec 7, 2023 · Event occurred Sep 22, 2022
Report 3003899624-2023-00006 · MDR key 18280667
Device
Generic name
104 Led Cluster Probe Wavelength = 660nm & 850nm Issue 1
Manufacturer
Thor Photomedicine LtdModel number
S2140Product problems
- Excessive Heating
- Excessive Heating
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
FAULT CONFIRMED - DIODES REPLACED. THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORT BASED ON ENHANCEMENTS MADE TO THE COMPANY'S COMPLAINT HANDLING PROCESSES. THIS EVENT IS BEING FILED IN ACCORDANCE WITH A CAPA WHICH HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.
Description of Event or Problem
PROBE BECOMES VERY HOT.
Remedial action
- Repair