69 LED CLUSTER PROBE WAVELENGTH = 660NM & 850NM ISSUE 1
Received Dec 7, 2023 · Event occurred May 20, 2021
Report 3003899624-2023-00010 · MDR key 18280665
Device
Generic name
69 Led Cluster Probe Wavelength = 660nm & 850nm Issue 1
Manufacturer
Thor Photomedicine LtdModel number
S2110Product problems
- Excessive Heating
- Excessive Heating
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE RINGS AROUND THE PROBES GET VERY HOT ESPECIALLY WHEN DOING THE SMALL PROBE ON STEM CELLS AND IT CONCERNS ME THAT IT SHOULD NOT BE GETTING THIS HOT.
Additional Manufacturer Narrative
TEMPERATURE TESTING COMPLETED FOR 69 CLUSTER AND RESULTS WITHIN SPECIFICATION. NO ACTION REQUIRED. THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORT BASED ON ENHANCEMENTS MADE TO THE COMPANY'S COMPLAINT HANDLING PROCESSES. THIS EVENT IS BEING FILED IN ACCORDANCE WITH A CAPA WHICH HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.
Remedial action
- Inspection