104 LED CLUSTER PROBE WAVELENGTH = 660NM & 850NM ISS 1

Received Dec 6, 2023 · Event occurred Jan 10, 2023

Report 3003899614-2023-00005 · MDR key 18279533

Device

Generic name

104 Led Cluster Probe Wavelength = 660nm & 850nm Iss 1

Model number

S2140

Product problems

  • Excessive Heating
  • Excessive Heating

Patient

Unknown

  • Burning Sensation
  • Burning Sensation

Narrative

Additional Manufacturer Narrative

THE PROBE HEAD REQUIRED CLEANING. AFTER CLEANING, PROBE WORKED WITHIN SPECIFICATIONS. THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORT BASED ON ENHANCEMENTS MADE TO THE COMPANY'S COMPLAINT HANDLING PROCESSES. THIS EVENT IS BEING FILED IN ACCORDANCE WITH A CAPA WHICH HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.

Description of Event or Problem

LED CLUSTER PROBE IS GETTING TOO HOT TO USE.