inforMED
MalfunctionILY

810NM 1W CLUSTER (5X200MW) ISSUE 1

Received Dec 6, 2023 · Event occurred Jun 5, 2023

Report 3003899624-2023-00004 · MDR key 18279422

Device

Generic name

810nm 1w Cluster (5x200mw) Issue 1

Model number

S2160

Product problems

  • Excessive Heating
  • Excessive Heating

Patient

Unknown

  • Burning Sensation
  • Burning Sensation

Narrative

Additional Manufacturer Narrative

FAULTS CONFIRMED. BLUE WIRE SHORTED TO GROUND IN 1.5M CABLE. REPLACED COMPLETE CABLE. NO CORRECTIVE ACTION WAS REQUIRED. THIS MDR IS BEING SUBMITTED AS A PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORT BASED ON ENHANCEMENTS MADE TO THE COMPANY'S COMPLAINT HANDLING PROCESSES. THIS EVENT IS BEING FILED IN ACCORDANCE WITH A CAPA WHICH HAS BEEN OPENED TO MANAGE THE ACTIONS RELATED TO REMEDIATION OF COMPLAINT FILES AND ANY REQUIRED MDR REPORTING.

Description of Event or Problem

ON THE LAST USE, THE MIDDLE SECTION OF THE 810NM 1W CLUSTER (5X200MW) ISSUE 1 WAS HOT TO TOUCH - NOT THE HEAD BUT THE MID SECTION OF THE LEAD.

Remedial action

  • Repair