MalfunctionHRT

QUAD CUTTER

Received Nov 28, 2023

Report 1219602-2023-02318 · MDR key 18222033

Device

Generic name

Stripper, Surgical

Catalog number

72205610

Lot number

51072413

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

H10: INTERNAL COMPLAINT REFERENCE: (B)(4).

Description of Event or Problem

IT WAS REPORTED THAT THE CUTTER MECHANISM OF THE QUAD CUTTER BROKE OFF FROM THE SHAFT WHEN THE BLADE WAS BEING REMOVED. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT AND IF THERE WAS A BACK-UP DEVICE AVAILABLE OR IF THERE WAS A DELAY.

Additional Manufacturer Narrative

THE REPORTED DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT IT IS NOT IN ITS ORIGINAL PACKAGING. THE OUTER BLADE, PROXIMAL CUTTER IS FRACTURED AWAY FROM THE SHAFT. THE EDGES ARE DEFORMED AND HAVE SLIVERS OF METAL SHORN AWAY. THERE ARE SEVERAL SCUFFS IN THE METAL FROM TOOTHED INSTRUMENTS. A REVIEW OF DEVICE RECORDS SHOWED THERE WERE NO INDICATIONS TO SUGGEST THAT THE PRODUCT DID NOT MEET MANUFACTURING SPECIFICATIONS UPON RELEASE FOR DISTRIBUTION. A COMPLAINT HISTORY REVIEW FOUND SIMILAR REPORTED EVENTS. A RISK MANAGEMENT REVIEW FOUND THAT THE REPORTED FAILURE AND/OR HARM WAS DOCUMENTED APPROPRIATELY, AND THERE WERE NO INDICATIONS TO SUGGEST THE ANTICIPATED RISK IS NOT ADEQUATE. THE ROOT CAUSE HAS BEEN ASSOCIATED WITH UNINTENDED USE OF THE DEVICE. FACTORS THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT INCLUDE EXCESSIVE FORCE ON THE DEVICE, ATTEMPTED CORRECTION OF A DAMAGED DEVICE, OR AN INADVERTENT IMPACT EVENT INCONSISTENT WITH NORMAL USE. NO CONTAINMENT OR CORRECTIVE ACTIONS ARE RECOMMENDED AT THIS TIME.