Received Nov 28, 2023 · Event occurred Nov 14, 2023

Report 9610877-2023-61271 · MDR key 18213898

Device

Generic name

Video Esophagoscope

Model number

EE17-J10

Product problems

  • Erratic or Intermittent Display
  • Erratic or Intermittent Display

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE.

Additional Manufacturer Narrative

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EE-1580K IS AVAILABLE IN THE USA WITH A 510K NUMBER K961564. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE IMAGE HORIZONTAL LINES. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE IMAGE. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.