AMSCO 600 STERILIZER
Received Nov 21, 2023 · Event occurred Oct 23, 2023
Report 3005899764-2023-00067 · MDR key 18183403
Device
Generic name
Amsco 600 Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Model number
AMSCO 600 STEAM STERILIZERCatalog number
EF1201110
Product problems
- Use of Device Problem
- Use of Device Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A STERIS SERVICE TECHNICIAN INSPECTED THE AMSCO 600 STERILIZER AND FOUND THE UNIT TO BE OPERATING PROPERLY. NO ISSUES WERE NOTED WITH THE FUNCTION OR OPERATION AND THE REPORTED EVENT WAS UNABLE TO BE DUPLICATED. THE TECHNICIAN STATED THAT USER FACILITY PERSONNEL WERE UTILIZING STERILIZATION TRAYS THAT WERE WEIGHING 37 POUNDS. THE MAXIMUM WEIGHT OF THE TRAYS SHOULD BE 25 POUNDS. IF THE TRAY IS TOO HEAVY, THE STERILIZER WILL HAVE DIFFICULTY PULLING THE MOISTURE OUT OF THE CHAMBER DURING A PROCESSING CYCLE. THE OPERATOR MANUAL (PG. 5) HAS A RECOMMENDED LOAD (MAXIMUM) OF 25 POUNDS FOR INSTRUMENT TRAYS FOR SPECIFIC CYCLE TYPES, STERILIZATION TEMPERATURES, STERILIZE TIME AND DRY TIME. THE TECHNICIAN ALSO OBSERVED THAT THE STEAM FILTER ASSEMBLY CONNECTED TO THE STERILIZER REQUIRED SERVICE. THE STEAM FILTER ASSEMBLY IS NOT A STERIS DEVICE AND IS SERVICED AND MAINTAINED BY THE USER FACILITY. THE USER FACILITY IS RESPONSIBLE FOR THE SERVICE AND MAINTENANCE OF THE STEAM FILTER ASSEMBLY. THE TECHNICIAN STATED THAT IF THE STEAM FILTERS ARE CLOGGED OR THE STEAM TRAP IS NOT FUNCTIONING PROPERLY, WATER CAN BUILD UP AND GET PULLED INTO THE STERILIZER'S CHAMBER. THE STERIS SERVICE TECHNICIAN DISCUSSED THE REPORTED EVENT WITH USER FACILITY PERSONNEL AND COUNSELED THEM ON THE PROPER USE AND OPERATION OF THE AMSCO 600 STERILIZER, SPECIFICALLY UTILIZING THE TRAY WEIGHT AND SERVICE NEEDED FOR THE STEAM FILTER ASSEMBLY. THE STERIS SERVICE TECHNICIAN RETURNED THE AMSCO 600 STERILIZER TO SERVICE AND NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Description of Event or Problem
THE USER FACILITY REPORTED THAT WATER WAS REMAINING IN THE INSTRUMENT TRAYS AFTER BEING PROCESSED IN THEIR AMSCO 600 STERILIZER. PROCEDURE DELAYS AND A PROCEDURE CANCELLATION WERE REPORTED. NO REPORT OF INJURY.