inforMED
MalfunctionMMX

CLOVERSNARE¿ 4-LOOP VASCULAR RETRIEVER

Received Nov 21, 2023

Report 1820334-2023-01605 · MDR key 18182291

Device

Generic name

Mmx Device, Percutaneous Retrieval

Manufacturer

Cook Inc

Catalog number

VRS-6.0-90

Lot number

15075886

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • Insufficient Information
  • No Clinical Signs, Symptoms or Conditions
  • Insufficient Information
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR UNCHANGED. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem

ADDITIONAL INFORMATION WAS RECEIVED 12DEC2023. THE TYPE OF FILTER BEING RETRIEVED AT THE TIME OF THE EVENT IS UNKNOWN, AND IT IS UNKNOWN HOW LONG THE FILTER HAD BEEN IN PLACE. THE FILTER FEET WERE NOT EMBEDDED IN THE CAVAL WALL AND THE FILTER WAS NOT DIFFICULT TO RETRIEVE. THE COAXIAL SHEATH SYSTEM WAS ADVANCED COMPLETELY OVER THE FILTER, WITH THE OUTER SHEATH ADVANCED PAST THE TIP OF THE INNER SHEATH, AND THE FILTER WAS FULLY COLLAPSED. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT.

Additional Manufacturer Narrative

SUMMARY OF EVENT: AS REPORTED, DURING RETRIEVAL OF AN UNKNOWN INFERIOR VENA CAVA FILTER, PART OF A CLOVERSNARE 4-LOOP VASCULAR RETRIEVER SET'S CATHETER BROKE OFF AND WAS STUCK TO THE FILTER. REPORTEDLY, THE CATHETER WAS INTACT UPON ADVANCEMENT INTO THE BODY. THE FILTER WAS RETRIEVED, AND THE FILTER AND RETRIEVAL DEVICE WERE REMOVED FROM THE PATIENT. UPON INSPECTION OF THE FILTER, THE USER NOTED THAT A PIECE OF THE CATHETER WAS BROKEN OFF AND STUCK TO THE FILTER. THE USER DOES NOT BELIEVE THAT ANY PART OF THE DEVICE WAS LEFT IN THE PATIENT. PHOTOS PROVIDED BY THE CUSTOMER SHOW DAMAGE TO THE DISTAL TIP OF THE INNER COAXIAL SHEATH, WITH POSSIBLE SEPARATION OF MATERIAL. ADDITIONAL INFORMATION WAS RECEIVED 12DEC2023. THE TYPE OF FILTER BEING RETRIEVED AT THE TIME OF THE EVENT IS UNKNOWN, AND IT IS UNKNOWN HOW LONG THE FILTER HAD BEEN IN PLACE. THE FILTER FEET WERE NOT EMBEDDED IN THE CAVAL WALL, AND THE FILTER WAS NOT DIFFICULT TO RETRIEVE. THE COAXIAL SHEATH SYSTEM WAS ADVANCED COMPLETELY OVER THE FILTER, WITH THE OUTER SHEATH ADVANCED PAST THE TIP OF THE INNER SHEATH, AND THE FILTER WAS FULLY COLLAPSED. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT. INVESTIGATION EVALUATION: REVIEWS OF THE COMPLAINT HISTORY, DEVICE HISTORY RECORD (DHR), INSTRUCTIONS FOR USE (IFU), MANUFACTURING INSTRUCTIONS, AND QUALITY CONTROL PROCEDURES WERE CONDUCTED DURING THE INVESTIGATION. A VISUAL INSPECTION OF DEVICE PHOTOS, PROVIDED BY THE CUSTOMER, WAS ALSO CONDUCTED. THE COMPLAINT DEVICE WAS NOT RETURNED TO COOK FOR INVESTIGATION; HOWEVER, PHOTOS WERE PROVIDED. THE DISTAL TIP OF THE INNER COAXIAL SHEATH IS DAMAGED, WITH POSSIBLE SEPARATION OF MATERIAL. THE SNARE CATHETER TIP DOES NOT APPEAR TO BE DAMAGED OR SEPARATED. IT IS LIKELY THAT THE SEPARATED FRAGMENT WAS PART OF THE INNER SHEATH¿S TIP. A DOCUMENT-BASED INVESTIGATION EVALUATION WAS PERFORMED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO RELEVANT NON-CONFORMANCES ON THE LOT. A REVIEW OF COMPLAINT HISTORY FOUND NO ADDITIONAL COMPLAINTS FOR THIS LOT NUMBER. THE PRODUCT IFU STATES "WHILE HOLDING THE CLEAR Y-FITTING STEADY, ADVANCE THE COAXIAL SHEATH SYSTEM OVER THE FOREIGN OBJECT. NOTE: THE OUTER SHEATH OF THE COAXIAL SYSTEM MAY BE ADVANCED OVER THE TIP OF THE INNER SHEATH TO COVER ANY PORTION OF THE FOREIGN OBJECT NOT CONTAINED INSIDE THE DISTAL TIP OF THE INNER SHEATH." A REVIEW OF THE DEVICE MASTER RECORD (DMR) CONCLUDED THAT SUFFICIENT INSPECTION ACTIVITIES ARE IN PLACE TO IDENTIFY THIS FAILURE MODE PRIOR TO DISTRIBUTION. THE INFORMATION PROVIDED UPON REVIEW OF THE DMR, DHR, IFU, AND INVESTIGATION OF DEVICE PHOTOS SUGGESTS THAT THERE IS EVIDENCE THE DEVICE WAS MANUFACTURED TO SPECIFICATION. THERE IS NO EVIDENCE OF NON-CONFORMING DEVICES IN-HOUSE OR IN THE FIELD. BASED ON THE INFORMATION PROVIDED AND THE RESULTS OF THE INVESTIGATION, COOK HAS CONCLUDED THAT A COMPONENT FAILURE, UNRELATED TO MANUFACTURING OR DESIGN DEFICIENCIES, CONTRIBUTED TO THIS EVENT. THE EXACT CONDITIONS EXPERIENCED DURING THE PROCEDURE CANNOT BE REPLICATED. THE RISK ANALYSIS FOR THIS FAILURE MODE WAS REVIEWED AND NO ADDITIONAL ESCALATION WAS REQUIRED. THE APPROPRIATE PERSONNEL HAVE BEEN NOTIFIED AND COOK WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Additional Manufacturer Narrative

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. D9, H3: IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED. E3: OCCUPATION = INVENTORY CONTROL ANALYST. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem

AS REPORTED, DURING RETRIEVAL OF AN UNKNOWN INFERIOR VENA CAVA FILTER, PART OF A CLOVERSNARE 4-LOOP VASCULAR RETRIEVER SET'S CATHETER BROKE OFF AND WAS STUCK TO THE FILTER. REPORTEDLY, THE CATHETER WAS INTACT UPON ADVANCEMENT INTO THE BODY. THE FILTER WAS RETRIEVED, AND THE FILTER AND RETRIEVAL DEVICE WERE REMOVED FROM THE PATIENT. UPON INSPECTION OF THE FILTER, THE USER NOTED THAT A PIECE OF THE CATHETER WAS BROKEN OFF AND STUCK TO THE FILTER. THE USER DOES NOT BELIEVE THAT ANY PART OF THE DEVICE WAS LEFT IN THE PATIENT. PHOTOS PROVIDED BY THE CUSTOMER SHOW DAMAGE TO THE DISTAL TIP OF THE INNER COAXIAL SHEATH, WITH POSSIBLE SEPARATION OF MATERIAL. ADDITIONAL INFORMATION HAS BEEN REQUESTED.