MalfunctionGKT

SPECTRA OPTIA®

Received Nov 16, 2023 · Event occurred Feb 14, 2023

Report 18147602 · MDR key 18147602

Device

Generic name

Separator, Automated, Blood Cell, Diagnostic

Manufacturer

Terumo Bct, Inc.

Lot number

2211216241

Patient

23360 DA · Male

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

PATIENT WAS UNDERGOING APHERESIS STEM CELL COLLECTION. HEALTHCARE PROVIDERS IDENTIFIED RETURN PRESSURE ALARMS WITH POOR FLOW. UNABLE TO OBTAIN BLOOD FLOW GREATER THAN 50-60ML/MIN. FLUSHING, REVERSING AND REPOSITIONING OF PATIENT WAS ATTEMPTED WITHOUT CHANGE TO. UPON INSPECTION OF TUBING DURING ATTEMPTS AT FIXING RETURN FLOW, IDENTIFIED HERMETIC SEAL ON RETURN LINE. NEW KIT SET-UP/PRIMED. NO OTHER EXPERIENCED ONCE KIT CHANGED.