inforMED
InjuryFTC

UV BIOTEK

Received Aug 19, 2010 · Event occurred Aug 17, 2010

Report MW5017185 · MDR key 1813796

Device

Generic name

Phototherapy Unit

Manufacturer

Uv Biotek

Model number

3200NB

Product problems

  • Output above Specifications
  • Environmental Compatibility Problem
  • Failure to Shut Off

Patient

8 YR

  • Radiation Burn
  • Erythema
  • Pain
  • Skin Discoloration
  • Superficial (First Degree) Burn

Narrative

Description of Event or Problem

THERE WAS A POWER OUTAGE IN THE BUILDING FROM (B)(6)2010. THE POWER WAS RESTORED ON (B)(6)2010 BY 11:30 OR SO PM. OURS IS A DERMATOLOGY PRACTICE. WE USE LIGHT THERAPY FOR VARIOUS DISORDERS. OUR LIGHT BOX IS MADE BY UV BIOTEK. WE HAVE USED IT FOR THE LAST FOUR YEARS WITH NO PROBLEMS. ON (B)(6)2010, MY STAFF PLACED AN (B)(6) CHILD IN THE LIGHT BOX FOR ONE MINUTE AND 15 SECONDS. THE LIGHT FAILED TO TURN OFF. A STAFF MEMBER REALIZED THIS AND TURNED THE LIGHT OFF MANUALLY. THE CHILD WAS GIVEN ABOUT A THREE MINUTE DOSE. NORMALLY THE SAFETY SHUT OFF SHOULD DELIVER ONLY THE AMOUNT OF LIGHT THAT IS PRE SET BY THE TIMER. FOR SOME REASON ON THIS DAY, THE LIGHT BOX FAILED TO SHUT DOWN AS IT USUALLY DOES. THIS IS HIGHLY UNUSUAL. WE SUSPECT AN ELECTRICAL PROBLEM. THE CHILD DEVELOPED MARKED ERYTHEMA ON THE FACE AND AREAS OF VITILIGO AND A SMALL IRREGULAR SHAPED PATCH OF SOLAR BURN ON THE LEFT BUTTOCK. BY (B)(6)2010, HE HAD PAIN ON THE BUTTOCK ONLY AND BLANCHING ERYTHEMA BUT NO TENDERNESS ELSEWHERE AND THE FACIAL ERYTHEMA HAD NEARLY RESOLVED. DATES OF USE: (B)(6)2006 - (B)(6)2010. DIAGNOSIS OR REASON FOR USE: VITILIGO, PSORIASIS.