MalfunctionLCP

TINA-QUANT HEMOGLOBIN A1CDX GEN. 3

Received Nov 6, 2023 · Event occurred Sep 30, 2023

Report 1823260-2023-03520 · MDR key 18072463

Device

Generic name

Glycosylated Hemoglobin Assay

Manufacturer

Roche Diagnostics

Catalog number

08056668190

Lot number

70882001

Product problems

  • High Test Results
  • High Test Results

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BASED ON THE INFORMATION PROVIDED, THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE CUSTOMER'S COBAS PRO C 503 ANALYTICAL UNIT IS 2058-03. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER RECEIVED QUESTIONABLE TINA-QUANT HEMOGLOBIN A1CDX GEN. 3 RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 7.65 %. THE REPEAT RESULT WAS 5.58%. THE REPEAT RESULT WAS BELIEVED TO BE THE TRUE RESULT BASED ON THE PATIENT'S HISTORY AND THE GLUCOSE RESULT.