inforMED
MalfunctionJQW

BD FACS¿ LWA

Received Nov 3, 2023 · Event occurred Oct 23, 2023

Report 2916837-2023-00214 · MDR key 18066619

Device

Generic name

Station, Pipetting Diluting Clinical Use

Catalog number

337408

Lot number

NA

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE FOLLOWING FIELDS HAVE BEEN UPDATED: G.1. REPORTING OFFICE: BECTON DICKINSON AND COMPANY BD BIOSCIENCES G.2. REPORTING OFFICE CONTACT: (B)(6). INVESTIGATION SUMMARY: BASED ON THE REVIEW OF THE COMPLAINT TREND, DEFECT TREND, DHR, RISK ANALYSIS AND SERVICE ACTIVITY REVIEW THE COMPLAINT WAS CONFIRMED, AND THE POTENTIAL CAUSE OF THE UNEXPECTED RESULTS WAS DETERMINED TO BE WORN OUT CELL WASH ASSEMBLY AND V3. THE FSE REPLACED THE PARTS AND AFTER SUBSEQUENT TESTING THE INSTRUMENT WAS FUNCTIONING AS EXPECTED. NO ONE WAS HARMED OR INJURED, AND NO PATIENTS WERE DIAGNOSED OR TREATED BASED ON ANY UNEXPECTED RESULTS. THE SAFETY RISK OF THIS HAZARD HAS BEEN IDENTIFIED TO BE WITHIN THE ACCEPTABLE LEVEL. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem

IT WAS REPORTED THAT ERRONEOUS RESULTS WERE OBSERVED WHILE USING THE BD FACS¿ LWA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MDR SAFETY CHECKLIST - PATIENT SAMPLES (CASE) 1. ARE THERE ERRONEOUS RESULTS ON PATIENT SAMPLES FROM DIAGNOSTIC TEST? (IF YES OR UNKNOWN, GO TO QUESTION #2. IF NO, NO FURTHER QUESTIONS REQUIRED.): YES 2. WAS THERE ANY DELAY OF TREATMENT DUE TO THE ISSUE? NO 3. IF PATIENT SAMPLES WERE REDRAWN, WAS THERE ANY CHANGE OR DELAY OF TREATMENT? NOT APPLICABLE 4. WAS THERE ANY PHYSICAL HARM/INJURY TO THE PATIENT DUE TO THE ISSUE? NO 5. PROVIDE DETAILS - HOW AND TO WHAT EXTENT? THEY RERAN SAMPLE ON A DIFFERENT INSTRUMENT WHETHER CARRYOVER HAPPENED ON PATIENT SAMPLES? THERE WAS NO CARRY OVER IT WAS REPORTED BY THE CUSTOMER THAT SYSTEM IS HAVING PROBLEMS WASHING SAMPLES ON SPECIFIC PROGRAM.

Additional Manufacturer Narrative

D.4. MEDICAL DEVICE EXPIRATION DATE: NA. E.1. INITIAL REPORTER ADDRESS 1: (B)(6). H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED THAT ERRONEOUS RESULTS WERE OBSERVED WHILE USING THE BD FACS¿ LWA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MDR SAFETY CHECKLIST - PATIENT SAMPLES (CASE). 1. ARE THERE ERRONEOUS RESULTS ON PATIENT SAMPLES FROM DIAGNOSTIC TEST? (IF YES OR UNKNOWN, GO TO QUESTION #2. IF NO, NO FURTHER QUESTIONS REQUIRED.): YES. 2. WAS THERE ANY DELAY OF TREATMENT DUE TO THE ISSUE? NO. 3. IF PATIENT SAMPLES WERE REDRAWN, WAS THERE ANY CHANGE OR DELAY OF TREATMENT? NOT APPLICABLE. 4. WAS THERE ANY PHYSICAL HARM/INJURY TO THE PATIENT DUE TO THE ISSUE? NO. 5. PROVIDE DETAILS - HOW AND TO WHAT EXTENT? THEY RERAN SAMPLE ON A DIFFERENT INSTRUMENT. WHETHER CARRYOVER HAPPENED ON PATIENT SAMPLES? THERE WAS NO CARRY OVER. IT WAS REPORTED BY THE CUSTOMER THAT SYSTEM IS HAVING PROBLEMS WASHING SAMPLES ON SPECIFIC PROGRAM.