4.5 VA-LCP CURVED COND PL/12 HOLE/266/RT
Received Nov 1, 2023
Report 8030965-2023-13651 · MDR key 18047152
Device
Generic name
Implant, Fixation Device, Condylar Plate
Manufacturer
Synthes GmbhCatalog number
02.124.412
Lot number
89P7430
Product problems
- Break
- Adverse Event Without Identified Device or Use Problem
- Break
- Adverse Event Without Identified Device or Use Problem
Patient
65 YR · Female
- Arthritis
- Fall
- Bone Fracture(s)
- Nerve Damage
- Pain
- Arthritis
- Fall
- Bone Fracture(s)
- Nerve Damage
- Pain
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. D1, D2, D3, D4, G5 ¿ 510K: THIS REPORT IS FOR AN UNKNOWN PLATES: 4.5MM VA LCP CONDYLAR PLATE/UNKNOWN LOT. PART AND LOT NUMBERS ARE UNKNOWN; UDI NUMBER IS UNKNOWN. D9: COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER. H3, H4, H6: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT ON OR ABOUT (B)(6) 2021 THE PATIENT UNDERWENT AN OPEN REDUCTION AND INTERNAL FIXATION OF A RIGHT PERIPROSTHETIC DISTAL FEMUR FRACTURE AND FEMORAL SUPRACONDYLAR FRACTURE, WHICH INVOLVED THE USE AND IMPLANTATION OF A 4.5MM VARIABLE ANGLE DISTAL FEMUR LIMITED CONTACT CONDYLAR PLATE AND SCREWS WHICH SUBSEQUENTLY MALFUNCTIONED OR FAILED ON OR ABOUT (B)(6) 2021 CAUSING INJURIES TO THE PATIENT. THE PATIENT SUSTAINED BODILY INJURIES, PAIN, SUFFERING AND SHOCK TO HER NERVES AND NERVOUS SYSTEM; WAS CAUSED TO BE INCAPACITATED FROM HER USUAL ACTIVITIES AND EMPLOYMENT. THIS REPORT IS FOR ONE (1) UNK - PLATES: 4.5MM VA LCP CONDYLAR PLATE. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Description of Event or Problem
THE BELOW COMPLAINTS ARE LINKED. THIS COMPLAINT WILL BE CAPTURED UNDER 2 SEPARATE COMPLAINTS AS LISTED BELOW: (B)(4).
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11: ADDITIONAL NARRATIVE: B5, D10 H3, H6: A PHOTO INVESTIGATION WAS COMPLETED: THE PHOTO/X-RAY INVESTIGATION REVEALED THAT THE DEVICE HAD SLIGHTLY DEFORMED ON THE DISTAL END OF THE PLATE. WITH THE SERIES OF X-RAYS THE BRAKEAGE OF THE IMPLANT IS NOT CLEAR, THEREFORE THE BROKEN ALLEGATION CANNOT BE CONFIRMED. WITH THE INFORMATION PROVIDED IS NOT POSSIBLE TO DETERMINE A POTENTIAL CAUSE AT THIS MOMENT, HOWEVER IN SUPPORT OF THE EVALUATION PERFORMED, THE OBSERVED DAMAGE OF THE DEVICE MAY HAVE BEEN CAUSED BY EXPOSURE TO IRREGULARITIES IN THE WEIGHT LOADING OF THE DEVICE AND MOMENTS GENERATED. THE MODE OF FAILURE OF THE DEVICE IS MULTI-FACTORIAL AND CONSIDERATION MUST BE GIVEN TO ALL OTHER POTENTIAL INFLUENCES SUCH AS SURGICAL PROCESS, PATIENT VARIABLES (E.G. ACTIVITY LEVEL AND USE), ANATOMICAL CONSIDERATIONS AND PATIENT CHANGES OVER TIME. ANY CONCLUSIONS FROM THE INVESTIGATIONAL INPUT PROVIDED HAVE TO BE PLACED INTO CONTEXT WITH ALL OTHER RELEVANT FACTORS. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS UNCONFIRMED AS THE OBSERVED CONDITION WOULD NOT CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
THIS IS REPORT IS 1 OF 1 FOR PC-(B)(4).
Description of Event or Problem
IT WAS FURTHER REPORTED: ON (B)(6) 2021 THE PATIENT PRESENTED TO THE EMERGENCY ROOM WITH COMPLAINT OF RIGHT LEG PAIN FOLLOWING A FALL. X-RAY SHOWED ACUTE ANGULATED FRACTURE INVOLVING THE DISTAL THIRD OF THE FEMORAL DIAPHYSIS TO INCLUDE A FRACTURE OF THE FIXATING METALLIC PLATE. ANGULATION AT THE FRACTURE SITE. ON (B)(6) 2021 THE PATIENT UNDERWENT RIGHT OPEN REDUCTION FEMUR FIXATION, INTRAMEDULLARY NAILING, AND REMOVAL OF HARDWARE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: G1. H4: PART NUMBER: 02.124.412. LOT NUMBER: 89P7430. MANUFACTURING SITE: MEZZOVICO. RELEASE TO WAREHOUSE DATE: 10 FEB 2021. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE NOT STERILE FINISHED LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.