LUCEA 10 RAIL
Received Oct 19, 2023 · Event occurred Oct 3, 2023
Report 9710055-2023-00782 · MDR key 17963757
Device
Generic name
Device, Medical Examination, Ac Powered
Manufacturer
Maquet SasModel number
ARDLCA409005ACatalog number
ARDLCA409005A
Product problems
- Crack
- Detachment of Device or Device Component
- Crack
- Detachment of Device or Device Component
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EXAMINATION LIGHTS - LUCEA 10 RAIL. IT WAS STATED THE HEADLIGHT COVER WAS CRACKED IN THE FIXING MADE BY 3 SCREWS WITH POSSIBILITY OF MISSING PARTICLES. ACCORDING TO THE TECHNICAL ANALYSIS, THE PROBLEM OCCURRED DUE TO IMPROPER USE OF THE DEVICE BY HOLDING THE DEVICE IN THE ILLUMINATION ZONE INSTEAD OF USING THE HANDLE. THE ISSUE WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION PROCEDURE MAY CAUSE CONTAMINATION. IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE EXAMINATION LIGHT DID NOT MEET ITS SPECIFICATION DUE TO CRACKS AND MISSING PARTICLES FROM COVER, WHICH CONTRIBUTED TO THE EVENT. . PROVIDED INFORMATION DOES NOT INDICATE IF UPON THE EVENT OCCURRENCE, THE DEVICE WAS OR WAS NOT BEING USED FOR PATIENT TREATMENT. REVIEW OF RECEIVED CUSTOMER PRODUCT COMPLAINTS RELATED TO INVESTIGATED ISSUE, REVEALED THAT THERE WERE NO INJURIES TO A USER NOR TO A PATIENT OR OPERATOR WHEN THIS PARTICULAR MALFUNCTION OCCURRED. WE CAN ASSUME THAT THE FAILURE RATIO OF CRACKS AND MISSING PARTICLES ON LUCEA 10/40 DEVICES IS MODERATE. A ROOT CAUSE ANALYSIS WAS PERFORMED BY SUBJECT MATTER EXPERTS AT MANUFACTURING SITE. BASED ON SOME INTERNAL TEST DONE BY MAQUET SAS (REF: CRE 12-085 FOR INSTANCE), ONLY ABNORMAL USE (VIOLENT COLLISIONS, EXCESSIVE PRESSURE¿) OR THE USE OF INCOMPATIBLE CLEANING PRODUCTS OR PROTOCOL, CAN DAMAGE THIS DEVICE AS REPORTED IN THIS COMPLAINT. TESTS PERFORMED BY GETINGE DID NOT LEAD TO CRACKS AS REPORTED IN THE COMPLAINT AT HAND. THE CRACKS DETECTED ON THE HANDLE INTERFACE AND TRANSPARENT COVER FIXING POINTS WERE PROBABLY CAUSED BY IMPROPER USE, COLLISIONS, INCOMPATIBLE CLEANING PRODUCTS OR PROTOCOL. TO PREVENT ANY INCIDENT THE LUCEA10/40 INSTRUCTION FOR USE 01701 EN12 MENTIONS TO CHECK THE INTEGRITY OF THE LIGHT HEADS DURING DAILY INSPECTION AND DESCRIBES HOW TO CLEAN AND DISINFECT THE LIGHT HEAD. THIS DOCUMENT INCLUDES SOME RECOMMENDED PRODUCTS AND SOME PROHIBITED PRODUCTS. IN ORDER TO AVOID MECHANICAL STRESSES APPLIED ON THE TRANSPARENT HOUSING DURING USE THE LUCEA10/40 INSTRUCTION FOR USE 01701 EN12 MENTIONS TO HANDLE THE LIGHT HEAD BY THE HANDLE. TO PREVENT ANY INCIDENT THE LUCEA10/40 INSTRUCTION FOR USE IFU 01701 MENTIONS TO CHECK THE INTEGRITY OF THE LIGHT HEADS DURING DAILY INSPECTION. FURTHERMORE THE LUCEA10/40 INSTRUCTION FOR USE IFU 01701 EN12 EXPLAINS HOW TO CLEAN AND DISINFECT THE LIGHT HEADS. THIS DOCUMENT INCLUDES SOME RECOMMENDED PRODUCTS AND SOME PROHIBITED PRODUCTS. IN ORDER TO AVOID TORQUES APPLIED ON THE TRANSPARENT HOUSING DURING USE THE LUCEA10/40 INSTRUCTION FOR USE IFU 01701 EN12 MENTIONS TO HANDLE THE LIGHT HEAD BY THE HANDLE (IFU_LUCEA_10_40_01701EN12, PAGES 22-25, 30). WE BELIEVE THE RELATED DEVICES ARE PERFORMING CORRECTLY IN THE MARKET. WE ALSO BELIEVE THAT IF THE MANUFACTURER RECOMMENDATION HAD BEEN FOLLOWED THE INCIDENT COULD HAVE BEEN AVOIDED. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME. THE CORRECTION OF H4 MANUFACTURE DATE FIELD DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS H4 MANUFACTURE DATE: 07/15/2021. CORRECTED H4 MANUFACTURE DATE: 07/16/2021.
Description of Event or Problem
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Description of Event or Problem
ON 9TH OCTOBER, 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EXAMINATION LIGHTS - LUCEA 10 RAIL. IT WAS STATED THE HEADLIGHT COVER WAS CRACKED IN THE FIXING MADE BY 3 SCREWS WITH POSSIBILITY OF MISSING PARTICLES. ACCORDING TO THE TECHNICAL ANALYSIS, THE PROBLEM OCCURRED DUE TO IMPROPER USE OF THE DEVICE BY HOLDING THE DEVICE IN THE ILLUMINATION ZONE INSTEAD OF USING THE HANDLE. THE ISSUE WAS CONFIRMED BY PHOTOGRAPHIC EVIDENCE. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTICLES FALLING OFF DURING EXAMINATION PROCEDURE MAY CAUSE CONTAMINATION.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. H3 OTHER TEXT : DEVICE NOT RETURNED TO MANUFACTURER.
Additional Manufacturer Narrative
THE CORRECTION OF B3 DATE OF EVENT FIELD DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS B3 DATE OF EVENT: 10/09/2023. CORRECTED B3 DATE OF EVENT: 10/03/2023.
Description of Event or Problem
MANUFACTURER'S REFERENCE NUMBER (B)(4).