inforMED
InjuryMMX

EN SNARE MINI

Received Oct 18, 2023 · Event occurred Oct 9, 2023

Report 9616662-2023-00020 · MDR key 17958413

Device

Generic name

Intravascular Snare

Catalog number

EN1003004/B

Lot number

K2607056

Product problems

  • Fracture
  • Fracture

Patient

Unknown

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Additional Manufacturer Narrative

THE SUSPECT DEVICE IS EXPECTED TO RETURN FOR EVALUATION. A FOLLOW UP WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem

CUSTOMER ALLEGES THE TIP OF THE SNARE BROKE OFF AND THEY USED ANOTHER SNARE TO REMOVE THE TIP. NO PATIENT INJURY.

Additional Manufacturer Narrative

THE SUSPECT DEVICE WAS RETURNED FOR EVALUATION. THE COMPLAINT WAS CONFIRMED. THE MOST PROBABLE CAUSE OF THIS OCCURRENCE IS FORCES IMPOSED UPON THE DEVICE WHEN TORTUOUS ANATOMY WAS ENCOUNTERED IN CLINIC. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND. THE DEVICE HISTORY RECORD WAS REVIEWED, AND NO EXCEPTION DOCUMENTS WERE FOUND.