SENSICA UO SYSTEM
Received Oct 3, 2023 · Event occurred Sep 12, 2023
Report 1018233-2023-07147 · MDR key 17860088
Device
Generic name
Sensica Uo System
Manufacturer
C.r. Bard, Inc. (covington) -1018233Catalog number
SCCS1002E
Lot number
NA
Product problems
- Defective Component
- Failure to Calibrate
- Defective Component
- Failure to Calibrate
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.
Description of Event or Problem
THE SENSICA URINE OUTPUT SYSTEM TOUCHSCREEN NEED UPDATE, AND FAILED THE DURING SCREEN CALIBRATION.
Additional Manufacturer Narrative
THE REPORTED ISSUE WAS CONFIRMED. THE ROOT CAUSE IDENTIFIED IS A FAILED TABLET/PC. THE DEVICE WAS EVALUATED UPON RECEIPT. THE SENSICA TOUCHSCREEN FAILED THE DURING CALIBRATION, AND TESTING. REPLACED TABLET/PC. OTHER MAINTENANCE PERFORMED. APPLIED THE OPTICAL SENSOR GASKET FOAM, THE TAMPER EVIDENT LABELS, AND NEW GROUND WIRE. COMPLETED FINAL FUNCTIONAL. DHR IS NOT REQUIRED AS THE REPORTED EVENT IS NOT AN OUT OF BOX FAILURE AND THEREFORE THE REPORTED EVENT IS NOT MANUFACTURING RELATED. THE LABELLING REVIEW SIS NOT REQUIRED AS THE COMPLAINT OR REPORTED ISSUE WAS CONFIRMED THROUGH OTHER ELEMENTS OF THE INVESTIGATION TO NOT BE LABELING OR PACKAGING RELATED. H11: SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE ACTUAL/SUSPECTED DEVICE WAS INSPECTED.
Description of Event or Problem
THE SENSICA URINE OUTPUT SYSTEM TOUCHSCREEN NEED UPDATE, AND FAILED THE DURING SCREEN CALIBRATION.