inforMED
MalfunctionGIO

BD SUREPREP¿ CAPILLARY TUBES- HEPARINIZED

Received Sep 29, 2023 · Event occurred Aug 31, 2023

Report 1119779-2023-01036 · MDR key 17845598

Device

Generic name

Tube, Collection, Capillary Blood

Catalog number

420317

Lot number

032823

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

MDR OWNERSHIP SHORT DESCRIPTION: NO EXACT EVENT TYPE IS AVAILABLE FOR BD SUREPREP¿ CAPILLARY TUBES LEAKING PATIENT SAMPLE, THEREFORE, LEAKING/BROKEN BACTEC BOTTLE - CONTAINING PATIENT SAMPLE HAS BEEN USED FOR REPORTABILITY PURPOSES. H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED THAT WHILE USING BD SUREPREP¿ CAPILLARY TUBES- HEPARINIZED, BLOOD IS SPILLING OUT OF TWO BOXES OF TUBES. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING BLOOD IS SPILLING OUT. HOW ARE THE CAPILLARY TUBES STORED WHEN NOT IN USE? TUBES ARE STORED AT ROOM TEMPERATURE IN ITS ORIGINAL PACKAGING. HOW ARE THE TUBES BEING FILLED? FILLED WITH WHOLE BLOOD BY RETRO ORBITAL BLEEDS AND SAPHENOUS VEIN. ONCE THE TUBE IS FILLED, HOW LONG IS THE BLOOD REMAINING IN CONTACT WITH THE SELF-SEALING PLUG? ONCE BLOOD HAS REACHED THE SELF-SEALING PLUG. THEN PLACED UPRIGHT PLUG FACING DOWNWARDS. QUANTITY OF TWO BOXES AFFECTED."

Additional Manufacturer Narrative

H.6. INVESTIGATION SUMMARY: COMPLAINT HISTORY REVIEW: A REVIEW OF PAST COMPLAINTS ON THIS PRODUCT OVER THE PAST 12 MONTHS DOES NOT INDICATE A TREND ON THIS ISSUE. DEVICE HISTORY RECORD REVIEW: THE REVIEW OF THE DEVICE HISTORY RECORD DID NOT INDICATE ANY MANUFACTURING ISSUES. SAMPLE ANALYSIS: A REVIEW OF THE PHOTOS PROVIDED DID NOT REVEAL ANY DEFECT. AN INSPECTION OF THE RETENTION SAMPLES WAS SATISFACTORY. NO DEFECTS WERE FOUND. EVALUATIONS RESULTS: BASED ON THE INVESTIGATION, NO DEFECT WAS OBSERVED. NO COMPLAINT TREND HAS BEEN OBSERVED. INVESTIGATION CONCLUSION: BASED ON THE EVALUATION OF THE INVESTIGATION, THE COMPLAINT WAS NOT CONFIRMED. NO FURTHER ACTIONS WILL BE TAKEN AS NO CONFIRMED TREND HAS BEEN IDENTIFIED. HOWEVER, BD WILL CONTINUE TO MONITOR FOR TRENDING. H3 OTHER TEXT : SEE H10.

Description of Event or Problem

IT WAS REPORTED THAT WHILE USING BD SUREPREP¿ CAPILLARY TUBES- HEPARINIZED, BLOOD IS SPILLING OUT OF TWO BOXES OF TUBES. NO PATIENT IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER IS REPORTING BLOOD IS SPILLING OUT. HOW ARE THE CAPILLARY TUBES STORED WHEN NOT IN USE? TUBES ARE STORED AT ROOM TEMPERATURE IN ITS ORIGINAL PACKAGING. HOW ARE THE TUBES BEING FILLED? FILLED WITH WHOLE BLOOD BY RETRO ORBITAL BLEEDS AND SAPHENOUS VEIN. ONCE THE TUBE IS FILLED, HOW LONG IS THE BLOOD REMAINING IN CONTACT WITH THE SELF-SEALING PLUG? ONCE BLOOD HAS REACHED THE SELF-SEALING PLUG. THEN PLACED UPRIGHT PLUG FACING DOWNWARDS. QUANTITY OF TWO BOXES AFFECTED.".