LAMINECT-PUNCH 40°ANGL W/4 L330
Received Sep 29, 2023 · Event occurred Aug 15, 2023
Report 8030965-2023-12362 · MDR key 17842882
Device
Product problems
- Contamination /Decontamination Problem
- Contamination /Decontamination Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. H3, H4, H6: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT ON (B)(6) 2023, THE DISTRIBUTOR REPORTED THAT A KERRISON RONGEUR (LOT NUMBER COULD NOT BE VERIFIED) HAD TISSUE BETWEEN ITS TWO MOVING ARMS TOWARDS THE MOUTH OF THE INSTRUMENT THAT COULD NOT BE CLEARED OR CLEANED. THE INSTRUMENT WAS FORWARDED ALONG TO ANOTHER DISTRIBUTOR WITH THIS SERVING AS A NOTICE THE INSTRUMENT NEEDS TO BE PROPERLY DISASSEMBLED AND CLEANED PRIOR TO ITS NEXT USE. NO FURTHER INCIDENTS HAVE BEEN REPORTED AND THE INSTRUMENT WILL NOT BE RETURNED. THERE IS NO KNOWN CLINICAL IMPACT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) LAMINECT-PUNCH 40°ANGL W/4 L330. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6 THE PRODUCT WAS NOT RETURNED TO DEPUY SYNTHES, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION DID NOT REVEAL ANY ISSUE WITH THE 389.410, LAMINECT-PUNCH 40°ANGL W/4 L330. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS UNCONFIRMED AS THE OBSERVED CONDITION OF THE 389.410, LAMINECT-PUNCH 40°ANGL W/4 L330 WOULD NOT CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.