QUATERA 700
Received Sep 20, 2023 · Event occurred Jul 3, 2023
Report 9615010-2023-00007 · MDR key 17786462
Device
Generic name
Ophthalmic, Anesthesia, Respiratory, Ent And Dental Devices Ophthalmic Device
Manufacturer
Carl Zeiss Meditec Ag (jena)Model number
QUATERA 700Catalog number
6407
Product problems
- Use of Device Problem
- Use of Device Problem
Patient
Unknown
- Eye Burn
- Eye Burn
Narrative
Additional Manufacturer Narrative
THE ROOT CAUSE OF THE REPORTED EVENT WAS AN EXCESSIVE ULTRASOUND POWER (U/S POWER) DUE TO USER ERROR. THE INSTRUCTION FOR USE STATES CLEARLY THE ACTIONS TO BE TAKEN TO AVOID APPLICATION OF EXCESSIVE U/S POWER.
Description of Event or Problem
CUSTOMER FROM GB (UNITED KINGDOM) INFORMED THAT A PATIENT CAME IN POST CATARACT SURGERY WITH A WHITE CORNEAL BURN. THE PATIENT UNDERWENT CATARACT SURGERY. THE WHITE CORNEAL BURN WAS DIAGNOSED IN A FOLLOW-UP CLINIC. THE PATIENT RECEIVED NACL 5% AND STEROID EYE DROPS FOR TREATMENT OF THE BURN. PATIENT HAS RECOVERED.