inforMED
MalfunctionFLE

AMSCO 400 STERILIZER

Received Sep 20, 2023 · Event occurred Aug 23, 2023

Report 3005899764-2023-00048 · MDR key 17782781

Device

Generic name

Sterilizer

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT STEAM WAS EMITTING FROM THEIR AMSCO 400 STERILIZER. NO REPORT OF INJURY.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN INSPECTED THE AMSCO 400 STERILIZER AND FOUND THAT ONE OF THE CONTACTORS ON THE GENERATOR'S HEATING ELEMENT WAS IN THE CLOSED POSITION. THE SAFETY VALVE RELEASED THE STEAM AS DESIGNED CAUSING THE REPORTED EVENT. THE TECHNICIAN REPLACED THE CONTACTOR AND HEATER ELEMENT, TESTED THE UNIT AND CONFIRMED IT TO BE OPERATING PROPERLY. THE AMSCO 400 STERILIZER WAS RETURNED TO SERVICE AND NO ADDITIONAL ISSUES HAVE BEEN REPORTED.