inforMED
MalfunctionHAM

CONMED

Received Jul 14, 2010 · Event occurred May 28, 2010

Report 1720159-2010-00048 · MDR key 1760672

Device

Generic name

Argon Beam Coagulation Handpiece

Catalog number

130321

Lot number

0908133

Product problems

  • Self-Activation or Keying

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

CONMED ELECTROSURGERY RECEIVED ONE ABC TRIPLE OPTION HANDPIECE FOR EVALUATION. DURING OUR EVALUATION, THE SUSPECT DEVICE SELF ACTIVATED.

Additional Manufacturer Narrative

OUR INVESTIGATION TEAM FOUND THE COAG BUTTON TO SELF ACTIVATE WHEN PLUGGED INTO THE ESU DURING TESTING. WE DISSECTED THE HANDPIECE AND FOUND NO IRREGULARITIES. OUR EVALUATION CONTINUED BY STRIPPING THE SUPPLY WIRE/CABLE TO PERFORM CONTINUITY TESTING. IT WAS OBSERVED, THE RED WIRE AND THE BLACK WIRE OF THE SUPPLY WIRE/CABLE TO BE SHORTED CAUSING THE SELF ACTIVATION. THE REPORTED MALFUNCTION WAS CONFIRMED. MANUFACTURING FACILITY WAS MADE AWARE REGARDING THIS COMPLAINT. OUR MANUFACTURING FACILITY WAS MADE AWARE OF THIS CONFIRMED EVENT. A DEVICE HISTORY RECORD REVIEW WAS PERFORMED WHICH REVEALED NO NONCONFORMANCES.