INFANT HEEL WARMER SQUEEZE 4X4IN
Received Aug 23, 2023 · Event occurred Jul 26, 2023
Report 1423537-2023-00983 · MDR key 17604324
Device
Generic name
Infant Heel Warmer (chemical Heat Pack)
Manufacturer
Mo016 Mo-moberlyModel number
11460-010TCatalog number
11460-010T
Lot number
V3E024
Product problems
- Burst Container or Vessel
- Burst Container or Vessel
Patient
Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A SAMPLE WAS RECEIVED FOR EVALUATION AND THE DEFECT REPORTED WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A SMALL HOLE IN THE OUTER POUCH DUE TO IMPROPER MACHINE SETUP. DEVICE HISTORY RECORD REVIEW WAS COMPLETED ON THE REPORTED LOT V3E024, AND THE LOT WAS MANUFACTURED AND RELEASED IN COMPLIANCE WITH ALL REQUIREMENTS. CARDINAL HEALTH HAS MADE A BUSINESS DECISION TO CLOSE THE SITE WHICH MANUFACTURES PRODUCT 11460-010T AND TRANSITION TO A THIRD-PARTY MANUFACTURER.
Additional Manufacturer Narrative
THE COMPLAINT WAS FORWARDED TO THE MANUFACTURING FACILITY WHERE IT IS CURRENTLY STILL UNDER INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE THE RESULTS HAVE BEEN COMPLETED.
Description of Event or Problem
CUSTOMER REPORTED THAT ¿IN THE NICU, THIS PRODUCT IS EXPLODING WHEN ACTIVATED. IT EXPLODED ONTO SEVERAL STAFF MEMBERS, POTENTIALLY CAUSING HARM TO A FEW. THE PACK CONTENTS CAME IN CONTACT WITH RN¿S EYES. I HAVE SOME OF THE EXPLODED PRODUCT IN MY OFFICE. WE HAVE REMOVED THE LOT NUMBER FROM THE FLOORS AND PATIENT ROOMS. A PROVIDER ON OUR UNIT INSPECTED HER EYES. EYE WASH DONE. ALL STAFF MEMBERS APPEAR TO BE DOING FINE. NO RESIDUAL COMPLAINTS NOTED¿.