TREVO NXT 6MM X 37MM - US
Received Aug 22, 2023 · Event occurred Aug 3, 2023
Report 3012931345-2023-00194 · MDR key 17600495
Device
Generic name
Neurovascular Mechanical Thrombectomy Device For Acute Ischemic Stroke Treatment
Manufacturer
Stryker Neurovascular-utah-salt Lake CityCatalog number
90315
Lot number
0000173684
Product problems
- Fracture
- Fracture
Patient
Unknown
- Hemorrhage/Blood Loss/Bleeding
- Vasoconstriction
- Foreign Body In Patient
- Hemorrhage/Blood Loss/Bleeding
- Vasoconstriction
- Foreign Body In Patient
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DURING THE ACUTE ISCHEMIC STROKE PROCEDURE WHERE THE MCA (MIDDLE CEREBRAL ARTERY) WAS OCCLUDED, PHYSICIAN FELT RESISTANCE WHILE ADVANCING AND RETRACTING THE SUBJECT STENT RETRIEVER INSIDE THE MICROCATHETER SHAFT. WHILE RETRACTION UNDER ASPIRATION, THE SUBJECT STENT RETRIEVER DETACHED FROM THE DELIVERY WIRE. THE SUBJECT STENT RETRIEVER WAS KINKED AND NARROWED IN THE VESSEL. PHYSICIAN TRIED TO RETRIEVE THE SUBJECT STENT RETRIEVER BUT WAS LEFT INSIDE THE PATIENT'S ANATOMY. M1 VESSEL REMAINED OCCLUDED AND DYNA CT REVEALED A BLEED POST PROCEDURE. PHYSICIAN WAS NOT CERTAIN IF THE BLEEDING WAS CAUSED BY THE SUBJECT STENT RETRIEVER. VASOSPASM WAS ALSO NOTED DURING THE PROCEDURE. PATIENT'S ANATOMY WAS VERY TORTUOUS. PHYSICIAN GAVE INTEGRILIN AT THE END OF THE PROCEDURE. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY AND THE PATIENT IS STILL IN ICU. NO ADDITIONAL INFORMATION AVAILABLE.
Additional Manufacturer Narrative
D4 CATALOG # - CORRECTED FROM 91315 TO 90315. THERE ARE CONTROLS IN THE MANUFACTURING PROCESS TO ENSURE THE PRODUCT MET SPECIFICATIONS UPON RELEASE. THE SUBJECT DEVICE IS NOT AVAILABLE; THEREFORE, FUNCTIONAL TESTING AS WELL AS VISUAL TESTING CANNOT BE PERFORMED. THE REPORTED EVENT IS COVERED IN THE DEVICE DIRECTIONS FOR USE (DFU). AS WELL, THE RISK OF THE REPORTED EVENT IS DOCUMENTED IN THE RISK DOCUMENTATION AND THERE ARE CURRENT CONTROLS TO MITIGATE THE RISK OF THE AS REPORTED EVENT. THE REPORTED EVENT COULD NOT BE CONFIRMED AND IT CANNOT BE CONFIRMED THAT THE DEVICE MET SPECIFICATION, AS THE DEVICE WAS NOT RETURNED. IT WAS REPORTED THAT THE STENT RETRIEVER DETACHED FROM THE DELIVERY WIRE WHEN THE PHYSICIAN WAS RETRACTING IT UNDER ASPIRATION. THE STENT RETRIEVER WAS KINKED AND NARROWED IN THE VESSEL AND WAS LEFT IN THE M1. M1 VESSEL REMAINED OCCLUDED AND DYNA COMPUTED TOMOGRAPHY (CT) REVEALED A BLEED. AS PER THE ADDITIONAL INFORMATION, RESISTANCE WAS EXPERIENCED DURING ADVANCEMENT OR RETRACTION OF THE RETRIEVER, VASOSPASM WAS NOTED DURING THE PROCEDURE, THE PATIENTS ANATOMY WAS VERY TORTUOUS, RESISTANCE WAS NOTED ADVANCING STENT RETRIEVER TO MIDDLE CEREBRAL ARTERY (MCA) AND RESISTANCE NOTED AT RETRIEVAL. AN ATTEMPT WAS MADE TO WITHDRAW THE RETRIEVER WITH INTEGRATED CLOT BACK THROUGH THE MICROCATHETER/ ASPIRATION CATHETER AND THE RETRIEVER WAS ROTATED OR TORQUED. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON A REVIEW OF ALL AVAILABLE INFORMATION THERE ARE FACTORS WHICH MAY HAVE CAUSED OR CONTRIBUTED TO THE RETRIEVER CORE WIRE FRACTURING, NAMELY, RESISTANCE DURING ADVANCEMENT/RETRIEVAL, VASOSPASM, TORTUOUS ANATOMY, ATTEMPTING TO WITHDRAW THE RETRIEVER/CLOT BACK THROUGH THE MICROCATHETER AND ROTATION/TORQUEING OF THE RETRIEVER. IT IS PROBABLE THAT ONE, OR A COMBINATION OF THESE FACTORS CAUSED THE RETRIEVER CORE WIRE TO FRACTURE. AN ASSIGNABLE CAUSE OF PROCEDURAL FACTORS WILL BE ASSIGNED TO THE REPORTED RETRIEVER CORE WIRE BROKEN DURING USE, RETRIEVER SHAPED SECTION DAMAGE, UN-RETRIEVED DEVICE FRAGMENTS AND RETRIEVER DIFFICULT/UNABLE TO GO THROUGH CATHETER SHAFT, AS THE ISSUE IS ASSOCIATED WITH A PRODUCT THAT MEETS STRYKER DESIGN AND MANUFACTURE SPECIFICATIONS AND WAS USED IN ACCORDING WITH THE DFU BUT DUE TO PROCEDURAL AND/OR ANATOMICAL FACTORS DURING USE, THE PRODUCT PERFORMANCE WAS LIMITED. IT CANNOT BE DEFINITIVELY DETERMINED IF THE PATIENT HEMORRHAGE AND VASOSPASM WERE CAUSED BY THE SUBJECT DEVICE. THE REPORTED PATIENT HEMORRHAGE AND VASOSPASM ARE KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND/OR PATIENT CONDITION NOTED WITH THE DIRECTIONS FOR USE, PRODUCT LABELING AND/OR RISK DOCUMENTATION FILES, THEREFORE A ASSIGNABLE CAUSE OF ANTICIPATED PROCEDURAL COMPLICATION WILL BE ASSIGNED TO THE REPORTED PATIENT HEMORRHAGE, BLOOD LOSS WITHOUT SEQUELAE AND PATIENT VASOSPASM SERIOUS.
Additional Manufacturer Narrative
THERE ARE CONTROLS IN THE MANUFACTURING PROCESS TO ENSURE THE PRODUCT MET SPECIFICATIONS UPON RELEASE. DURING VISUAL/MICROSCOPIC INSPECTION, THE CORE WIRE WAS SEEN TO BE BROKEN AT THE DISTAL END OF THE DEVICE, WITH TWO BREAK POINTS IDENTIFIED: THE FIRST APPROXIMATELY 189.5CM FROM THE PROXIMAL END AND THE SECOND ONE APPROXIMATELY 193 CM FROM THE PROXIMAL END. THE POLYMER JACKET WAS REMOVED, AND THE BREAK POINTS WERE IMAGED. THE RETRIEVER SHAPED SECTION WAS NOT RETURNED. THE INSERTION TOOL WAS NOT RETURNED. FUNCTIONAL INSPECTION WAS NOT CARRIED OUT DUE TO THE DAMAGE NOTED TO THE DEVICE. THE REPORTED EVENT IS COVERED IN THE DEVICE DIRECTIONS FOR USE (DFU). AS WELL, THE RISK OF THE REPORTED EVENT IS DOCUMENTED IN THE RISK DOCUMENTATION AND THERE ARE CURRENT CONTROLS TO MITIGATE THE RISK OF THE AS REPORTED EVENT. THE REPORTED RETRIEVER CORE WIRE BROKEN DURING USE WAS CONFIRMED. THE REPORTED PATIENT HEMORRHAGE, BLOOD LOSS WITHOUT SEQUELAE AND PATIENT VASOSPASM SERIOUS COULD NOT BE CONFIRMED AS THE EVENT IS PROCEDURE/PATIENT RELATED. THE REPORTED EVENT UN-RETRIEVED DEVICE FRAGMENTS, RETRIEVER SHAPED SECTION DAMAGE AND RETRIEVER DIFFICULT/UNABLE TO GO THROUGH CATHETER SHAFT COULD NOT BE DUPLICATED DURING DEVICE ANALYSIS; HOWEVER, THE ANALYSIS RESULTS ARE CONSISTENT WITH THE REPORTED EVENT. IT WAS REPORTED THAT THE SUBJECT STENT RETRIEVER DETACHED FROM THE DELIVERY WIRE WHEN THE PHYSICIAN WAS RETRACTING IT UNDER ASPIRATION. THE SUBJECT STENT RETRIEVER WAS KINKED AND NARROWED IN THE VESSEL AND WAS LEFT IN THE M1. M1 VESSEL REMAINED OCCLUDED AND DYNA CT REVEALED A BLEED. AS PER THE ADDITIONAL INFORMATION, RESISTANCE WAS EXPERIENCED DURING ADVANCEMENT OR RETRACTION OF THE RETRIEVER, VASOSPASM WAS NOTED DURING THE PROCEDURE, THE PATIENTS ANATOMY WAS VERY TORTUOUS, RESISTANCE WAS NOTED ADVANCING SUBJECT STENT RETRIEVER TO MCA AND RESISTANCE NOTED AT RETRIEVAL. AN ATTEMPT WAS MADE TO WITHDRAW THE RETRIEVER WITH INTEGRATED CLOT BACK THROUGH THE MICROCATHETER/ ASPIRATION CATHETER AND THE RETRIEVER WAS ROTATED OR TORQUED. THE DEVICE WAS RETURNED FOR ANALYSIS AND THE CORE WIRE WAS CONFIRMED TO BE FRACTURED IN TWO LOCATIONS. THE RETRIEVER SHAPED SECTION WAS NOT RETURNED. BASED ON A REVIEW OF ALL AVAILABLE INFORMATION THERE ARE MULTIPLE FACTORS WHICH MAY HAVE CAUSED OR CONTRIBUTED TO THE RETRIEVER CORE WIRE FRACTURING, NAMELY, RESISTANCE DURING ADVANCEMENT/RETRIEVAL OF THE RETRIEVER, VASOSPASM, TORTUOUS ANATOMY, ATTEMPTING TO WITHDRAW THE RETRIEVER/CLOT BACK THROUGH THE MICROCATHETER AND ROTATION/TORQUEING OF THE RETRIEVER DURING RETRACTION. IT IS PROBABLE THAT ONE, OR A COMBINATION OF THESE FACTORS CAUSED THE RETRIEVER CORE WIRE TO FRACTURE AND MAY ALSO BE RESPONSIBLE FOR THE SUBSEQUENT HEMORRHAGE. WHILE THE ASSOCIATED PATIENT HARMS PATIENT HEMORRHAGE, BLOOD LOSS WITHOUT SEQUELAE AND PATIENT VASOSPASM SERIOUS ARE ANTICIPATED IN NATURE AND DOCUMENTED AS SO IN THE DEVICE DIRECTIONS FOR USE, IN THIS CASE IT IS PROBABLE THAT THE DIFFICULTIES EXPERIENCED DURING THE CLINICAL PROCEDURE MAY HAVE CAUSED OR CONTRIBUTED TO THE PATIENT HARMS. AN ASSIGNABLE CAUSE OF PROCEDURAL FACTORS WILL BE ASSIGNED TO THE AS REPORTED AS WELL AS THE ANALYZED EVENTS OF RETRIEVER CORE WIRE BROKEN DURING USE AND TO AS REPORTED EVENTS OF RETRIEVER SHAPED SECTION DAMAGE, UN-RETRIEVED DEVICE FRAGMENTS, PATIENT HEMORRHAGE, BLOOD LOSS WITHOUT SEQUELAE, RETRIEVER DIFFICULT/UNABLE TO GO THROUGH CATHETER SHAFT AND PATIENT VASOSPASM SERIOUS. AS THE ISSUE IS ASSOCIATED WITH A PRODUCT THAT MEETS STRYKER DESIGN AND MANUFACTURE SPECIFICATIONS AND WAS USED IN ACCORDING WITH THE DFU BUT DUE TO PROCEDURAL AND/OR ANATOMICAL FACTORS DURING USE, THE PRODUCT PERFORMANCE WAS LIMITED.
Description of Event or Problem
IT WAS REPORTED THAT DURING THE ACUTE ISCHEMIC STROKE PROCEDURE WHERE THE MCA (MIDDLE CEREBRAL ARTERY) WAS OCCLUDED, PHYSICIAN FELT RESISTANCE WHILE ADVANCING AND RETRACTING THE SUBJECT STENT RETRIEVER INSIDE THE MICROCATHETER SHAFT. WHILE RETRACTION UNDER ASPIRATION, THE SUBJECT STENT RETRIEVER DETACHED FROM THE DELIVERY WIRE. THE SUBJECT STENT RETRIEVER WAS KINKED AND NARROWED IN THE VESSEL. PHYSICIAN TRIED TO RETRIEVE THE SUBJECT STENT RETRIEVER BUT WAS LEFT INSIDE THE PATIENT'S ANATOMY. M1 VESSEL REMAINED OCCLUDED AND DYNA CT REVEALED A BLEED POST PROCEDURE. PHYSICIAN WAS NOT CERTAIN IF THE BLEEDING WAS CAUSED BY THE SUBJECT STENT RETRIEVER. VASOSPASM WAS ALSO NOTED DURING THE PROCEDURE. PATIENT'S ANATOMY WAS VERY TORTUOUS. PHYSICIAN GAVE INTEGRILIN AT THE END OF THE PROCEDURE. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY AND THE PATIENT IS STILL IN ICU. NO ADDITIONAL INFORMATION AVAILABLE.
Additional Manufacturer Narrative
H3 OTHER TEXT : THE DEVICE IS NOT AVAILABLE TO THE MANUFACTURER.
Description of Event or Problem
IT WAS REPORTED THAT DURING THE ACUTE ISCHEMIC STROKE PROCEDURE WHERE THE MCA (MIDDLE CEREBRAL ARTERY) WAS OCCLUDED, PHYSICIAN FELT RESISTANCE WHILE ADVANCING AND RETRACTING THE SUBJECT STENT RETRIEVER INSIDE THE MICROCATHETER SHAFT. WHILE RETRACTION UNDER ASPIRATION, THE SUBJECT STENT RETRIEVER DETACHED FROM THE DELIVERY WIRE. THE SUBJECT STENT RETRIEVER WAS KINKED AND NARROWED IN THE VESSEL. PHYSICIAN TRIED TO RETRIEVE THE SUBJECT STENT RETRIEVER BUT WAS LEFT INSIDE THE PATIENT'S ANATOMY. M1 VESSEL REMAINED OCCLUDED AND DYNA CT REVEALED A BLEED POST PROCEDURE. PHYSICIAN WAS NOT CERTAIN IF THE BLEEDING WAS CAUSED BY THE SUBJECT STENT RETRIEVER. VASOSPASM WAS ALSO NOTED DURING THE PROCEDURE. PATIENT'S ANATOMY WAS VERY TORTUOUS. PHYSICIAN GAVE INTEGRILIN AT THE END OF THE PROCEDURE. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY AND THE PATIENT IS STILL IN ICU. NO ADDITIONAL INFORMATION AVAILABLE.