inforMED
MalfunctionIOY

SUPPORT ARM 176

Received Aug 22, 2023 · Event occurred Aug 7, 2023

Report 8010042-2023-01619 · MDR key 17595107

Device

Generic name

Support, Arm

Model number

SUPPORT ARM 176

Catalog number

6405976

Product problems

  • Fracture
  • Fracture

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE VENTILATOR'S SUPPORT ARM WAS BROKEN. THERE WAS NO PATIENT HARM REPORTED. MANUFACTURER'S REF. #: (B)(4).

Description of Event or Problem

MANUFACTURER'S REF. #: (B)(4).

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE VENTILATOR'S SUPPORT ARM WAS BROKEN. IDENTIFICATION OF ISSUE HAS BEEN DONE BY ANALYZING PROBLEM DESCRIPTION, SERVICE REPORT, COMMUNICATION WITH FIELD SERVICE TECHNICIAN (FSE) AND PROVIDED PICTURE. IT WAS REVEALED THAT CLAMP ATTACHMENT OF SUPPORT ARM WAS DEFECTIVE. ACCORDING TO THE INFORMATION PROVIDED BY THE TECHNICIAN, THE CUSTOMER BOUGHT A NEW SUPPORT ARM TO RESOLVE THE PROBLEM. THE ISSUE WAS DECIDED TO BE REPORTED AS THE BROKEN/DAMAGED SUPPORT ARM MAY LEAD TO STOP OF VENTILATION (EXTUBATION) OR INJURY. THERE WAS NO PATIENT HARM. THE MOST LIKELY ROOT CAUSE IS RELATED TO USABILITY AND TO DESIGN LEVEL AT THE SUPPLIER. H3 OTHER TEXT : 4112.