Received Aug 22, 2023 · Event occurred Aug 1, 2023

Report 1211566-2023-00010 · MDR key 17594878

Device

Generic name

Snarecoil Bone Biopsy Needle 14 Gauge

Manufacturer

Ranfac Corp

Model number

74050-01M

Catalog number

74050-01M

Lot number

49088

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

UPON INVESTIGATION, IT WAS NOTED THAT THE DEVICE WAS ACTUALLY AN RBN-133. BASED ON THE NOTE PROVIDED BY UPMC, IT IS ASSUMED THAT THE PROVIDER MIXED UP THE DEVICES. THERE WAS NO SAMPLE LOCATED WITHIN THE DEVICE AND WAS BLOCKED BY CORTICAL BONE. THE CORTICAL BONE WAS VISIBLE FROM THE DISTAL END OF THE DEVICE. CORTICAL BONE ENTERED THE LUMEN, BLOCKING THE NEEDLE FROM BEING ABLE TO RETRIEVE THE SPECIMEN. THIS IS TYPICALLY CAUSED BY PREMATURE REMOVAL OF THE STYLET DURING THE PROCEDURE.

Description of Event or Problem

1. RANFAC SNARECOIL BONE MARROW BIOPSY NEEDLE WHILE USING BIOPSY SAMPLE STUCK IN NEEDLE UNABLE TO DISLODGE FOR TESTING. HAD TO ATTEMPT FOR ANOTHER SAMPLE USING A DIFFERENT NEEDLE.