BD BBL¿ BRAIN HEART INFUSION BROTH
Received Aug 18, 2023 · Event occurred Aug 3, 2023
Report 1119779-2023-00904 · MDR key 17574023
Device
Generic name
Culture Media, General Nutrient Broth
Manufacturer
Becton, Dickinson & Co. (sparks)Catalog number
221788
Lot number
2279225
Product problems
- Contamination
- Contamination
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H.6. INVESTIGATION SUMMARY: MATERIAL 221788 IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER, AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE TUBES ARE FILLED, CAPPED, TORQUED, AND THEN LABELED BY MACHINE PER STANDARD OPERATING PROCEDURE (SOP). THE TUBES ARE TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, TUBES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE BATCH HISTORY RECORD REVIEW FOR BATCH 2279225 WAS SATISFACTORY PER INTERNAL PROCEDURES. FORMULATION, FILLING, TORQUING, AND AUTOCLAVING PROCESSES WERE WITHIN SPECIFICATIONS. IN PROCESS CHECKS WERE PERFORMED AT THE DESIGNATED INTERVALS. THOSE CHECKS CONFIRMED THAT THE CAPS WERE TIGHTENED TO THE VALIDATED SPECIFICATIONS PER INTERNAL PROCEDURE. QC INSPECTION AND TESTING WERE SATISFACTORY AT TIME OF RELEASE. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE REVIEW OF ITS COLOR AND CLARITY. SAMPLES SUBMITTED ARE EXAMINED TO ENSURE THAT THEY CONFORM TO TYPICAL LEVELS. THE APPEARANCE OF THIS BATCH WAS SATISFACTORY PER INTERNAL PROCEDURES. DIRECT STAINING TECHNIQUES ARE NOT PART OF QC RELEASE TESTING FOR THIS PRODUCT. AS PART OF THE RELEASE CRITERIA FOR THIS PRODUCT, THE BHR IS REVIEWED TO CONFIRM THE FOLLOWING: --THE TOTAL ELAPSED TIME BETWEEN END OF FORMULATION AND START OF THE AUTOCLAVE CYCLE WAS WITHIN THE SPECIFIED LIMITS. --ALL AUTOCLAVE PARAMETERS CONFORMED TO THE VALIDATED CYCLE PARAMETERS FOR THIS PRODUCT. --THE MINIMUM F0 FOR THIS PRODUCT WAS MET. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON THIS BATCH. RETENTION SAMPLES FROM BATCH 2279225 (10 TUBES) WERE AVAILABLE FOR INSPECTION. NO MEDIA DEFECTS WERE OBSERVED IN 10/10 RETENTION SAMPLES. ALL RETENTIONS TUBES HAD THE EXPECTED APPEARANCE FOR THIS PRODUCT OF LIGHT TO MEDIUM LIGHT YELLOW, TRACE HAZY TO CLEAR. FOR INVESTIGATION, TWO RETENTION TUBES WENT INTO INCUBATION. ONE RETENTION TUBE WAS PLACED INTO THE 20¿25-DEGREE CELSIUS INCUBATOR AND ONE RETENTION TUBE WAS PLACED IN THE 33¿37-DEGREE CELSIUS INCUBATOR. AT THE SEVENTH DAY OF INCUBATION THERE WERE NO SIGNS OF GROWTH OR TURBIDITY. THERE WAS NO CHANGE IN THE COLOR AND CLARITY IN OF THE MEDIA. THE MEDIA REMAINED MEDIUM YELLOW CLEAR TO TRACE HAZY AS DESCRIBED IN THE CERTIFICATE OF ANALYSIS. NO PHOTOS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION. NO RETURNS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION. THIS COMPLAINT CANNOT BE CONFIRMED. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION/APPEARANCE DEFECTS. THE COLOR AND CLARITY OF THE MEDIA ON BATCH 2279225 IS LIGHT TO MEDIUM LIGHT YELLOW, TRACE HAZY TO CLEAR WHICH IS DESCRIBED IN THE CERTIFICATE OF ANALYSIS. AS STATED IN THE IFU FOR MATERIAL 221788 (AVAILABLE ON BD.COM/E-LABELING), THIS MEDIA SHOULD BE REDUCED BY BOILING PRIOR TO USE. H3 OTHER TEXT : SEE H.10
Description of Event or Problem
IT WAS REPORTED THAT WHILE USING BD BBL¿ BRAIN HEART INFUSION BROTH CONTAMINATION WAS FOUND. NO PATIENT IMPACT OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PRESENCE OF "FLAKES" AND CLOUDINESS INSIDE THE TUBES WHEN DID THE INCIDENT OCCUR? DURING USE WE INOCULATED 5 BROTHS ON PVX AGAR PLATES BY FLOODING ON THE AFTERNOON OF THURSDAY 3 AUGUST AND THERE WAS A GROWTH.
Additional Manufacturer Narrative
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
Description of Event or Problem
IT WAS REPORTED THAT WHILE USING BD BBL¿ BRAIN HEART INFUSION BROTH CONTAMINATION WAS FOUND. NO PATIENT IMPACT OCCURRED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PRESENCE OF "FLAKES" AND CLOUDINESS INSIDE THE TUBES. WHEN DID THE INCIDENT OCCUR? DURING USE WE INOCULATED 5 BROTHS ON PVX AGAR PLATES BY FLOODING ON THE AFTERNOON OF THURSDAY 3 AUGUST AND THERE WAS A GROWTH.
Additional Manufacturer Narrative
THE FOLLOWING FIELDS HAVE BEEN UPDATED: B5. IT WAS REPORTED WHILE USING BD BBL¿ BRAIN HEART INFUSION BROTH, A TUBE WAS CONTAMINATED. THERE WAS NO REPORT OF PATIENT IMPACT. THIS SUPPLEMENTAL MDR IS BEING SUBMITTED AS PART OF A RETROSPECTIVE REVIEW OF RECORDS DATING APRIL 2023¿2025, UNDER CAPA 11910483.
Description of Event or Problem
IT WAS REPORTED WHILE USING BD BBL¿ BRAIN HEART INFUSION BROTH, A TUBE WAS CONTAMINATED. THERE WAS NO REPORT OF PATIENT IMPACT.