inforMED
MalfunctionMMX

CLOVERSNARE¿ 4-LOOP VASCULAR RETRIEVER

Received Aug 11, 2023 · Event occurred Jul 24, 2023

Report 1820334-2023-01075 · MDR key 17513330

Device

Generic name

Mmx Device, Percutaneous Retrieval

Manufacturer

Cook Inc

Catalog number

VRS-6.0-90

Lot number

15370648

Product problems

  • Break
  • Break

Patient

Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNCHANGED, OR UNAVAILABLE. SUMMARY OF EVENT: AS REPORTED, DURING REMOVAL OF AN UNSPECIFIED INFERIOR VENA CAVA FILTER, A CLOVERSNARE 4-LOOP VASCULAR RETRIEVER'S SHEATH STRETCHED AND SEPARATED AT THE HUB. INVESTIGATION EVALUATION: REVIEWS OF THE COMPLAINT HISTORY, DEVICE HISTORY RECORD, INSTRUCTIONS FOR USE (IFU), AND QUALITY CONTROL PROCEDURES WERE CONDUCTED DURING THE INVESTIGATION. THE COMPLAINT DEVICE WAS NOT RETURNED TO COOK FOR INVESTIGATION. A DOCUMENT-BASED INVESTIGATION EVALUATION WAS PERFORMED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO RELEVANT NON-CONFORMANCES ON THE LOT. A REVIEW OF COMPLAINT HISTORY FOUND NO ADDITIONAL COMPLAINTS FOR THIS LOT NUMBER. THE PRODUCT IFU STATES ¿UPON REMOVAL FROM PACKAGE, INSPECT THE PRODUCT TO ENSURE NO DAMAGE HAS OCCURRED.¿ THE INFORMATION PROVIDED UPON REVIEW OF THE DMR, DHR, AND IFU SUGGESTS THAT THERE IS EVIDENCE THE DEVICE WAS MANUFACTURED TO SPECIFICATION. THERE IS NO EVIDENCE OF NON-CONFORMING DEVICES IN-HOUSE OR IN THE FIELD. A REVIEW OF THE DEVICE MASTER RECORD (DMR) CONCLUDED THAT SUFFICIENT INSPECTION ACTIVITIES ARE IN PLACE TO IDENTIFY THIS FAILURE MODE PRIOR TO DISTRIBUTION. BASED ON THE INFORMATION PROVIDED AND THE RESULTS OF THE INVESTIGATION, COOK HAS CONCLUDED THAT A COMPONENT FAILURE, UNRELATED TO MANUFACTURING OR DESIGN DEFICIENCIES, CONTRIBUTED TO THIS EVENT. THE RISK ANALYSIS FOR THIS FAILURE MODE WAS REVIEWED AND NO ADDITIONAL ESCALATION WAS REQUIRED. THE APPROPRIATE PERSONNEL HAVE BEEN NOTIFIED AND COOK WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Description of Event or Problem

AS REPORTED, DURING REMOVAL OF AN UNSPECIFIED INFERIOR VENA CAVA FILTER, A CLOVERSNARE 4-LOOP VASCULAR RETRIEVER'S SHEATH STRETCHED AND SEPARATED AT THE HUB. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Additional Manufacturer Narrative

THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.