inforMED
MalfunctionCDP

ELECSYS FT3 III

Received Aug 7, 2023 · Event occurred Jun 26, 2023

Report 1823260-2023-02549 · MDR key 17473330

Device

Generic name

Total Triiodothyronine

Manufacturer

Roche Diagnostics

Catalog number

09005811190

Lot number

ASKU

Product problems

  • Non Reproducible Results
  • Non Reproducible Results

Patient

15 YR · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE PATIENT SAMPLE WAS RECEIVED FOR INVESTIGATION. THE INVESTIGATION WAS ABLE TO REPRODUCE THE CUSTOMER'S RESULTS. THE PATIENT SAMPLE WAS ANALYZED FOR INTERFERENCE AGAINST THE RUTHENIUM (RU) AND STREPTAVIDIN (SA) LABELS OF THE ASSAY. THERE WERE NO ASSAY-INTERFERING FACTORS IDENTIFIED. PRODUCT LABELING STATES "FOR DIAGNOSTIC PURPOSES, THE RESULTS SHOULD ALWAYS BE ASSESSED IN CONJUNCTION WITH THE PATIENT¿S MEDICAL HISTORY, CLINICAL EXAMINATION AND OTHER FINDINGS." THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Additional Manufacturer Narrative

THE FT3 III V2 REAGENT LOT NUMBERS USED AT THE CUSTOMER AND INVESTIGATION SITE'S E 801 WERE REQUESTED BUT NOT PROVIDED. THE FT3 III V2 REAGENT LOT NUMBER USED AT THE INVESTIGATION SITE'S E 411 IS 686656 WITH AN EXPIRATION DATE OF 01-APR-2024. THE PATIENT SAMPLE WAS REQUESTED FOR INVESTIGATION. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS FT3 III VER.2, ELECSYS FT4 III, AND ELECSYS FT4 IV RESULTS FROM ONE PATIENT SAMPLE TESTED ON THE CUSTOMER'S COBAS E 801 MODULE WITH SERIAL NUMBER (B)(6), THE INVESTIGATION SITE'S COBAS E 801 MODULE WITH UNKNOWN SERIAL NUMBER AND THE INVESTIGATION SITE'S COBAS E 411 MODULE WITH SERIAL NUMBER (B)(6). THIS MDR IS FOR THE FT3 III V2 ASSAY. REFER TO: MEDWATCH WITH A1 - PATIENT IDENTIFIER (B)(6) FOR THE FT4 IV ASSAY. MEDWATCH WITH A1 - PATIENT IDENTIFIER (B)(6) FOR THE FT4 III ASSAY. THE DATE OF BLOOD SAMPLING WAS USED AS THE DATE OF THE EVENT. THE INITIAL RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. THE REPORTER SUSPECTED SOME NONSPECIFIC INTERFERENCE IN THE ROCHE ASSAYS WHEN THEY COMPARED THE INITIAL RESULTS FROM THE E 801 TO THE REPEAT RESULTS OF THE PATIENT SAMPLE FROM THEIR WAKO ACCURASEED AND ABBOTT ALINITY ANALYZERS. THE PATIENT SAMPLE WAS THEN SENT TO THE INVESTIGATION SITE THAT USES A COBAS E 801 AND AN E 411. PLEASE REFER TO THE MEDWATCH (B)(6) FOR THE HIGHLIGHTED QUESTIONABLE RESULTS.