INFANT HEEL WARMER SQUEEZE 4X4IN
Received Jul 25, 2023 · Event occurred Jun 29, 2023
Report 1423537-2023-00969 · MDR key 17393895
Device
Generic name
Infant Heel Warmer (chemical Heat Pack)
Manufacturer
Mo016 Mo-moberlyModel number
11460-010TCatalog number
11460-010T
Lot number
V3D237
Product problems
- Burst Container or Vessel
- Burst Container or Vessel
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
CUSTOMER REPORTED, STAFF MEMBER SQUEEZED NEWBORN HEEL WARMER TO ACTIVATE AND THE GEL LEAKED OUT. IMMEDIATELY UPON SQUEEZING, THE PRODUCT PACKAGING SEPARATED ALONG AN EDGE CAUSING A HOLE AND INNER CONTENTS TO LEAK OUT. CUSTOMER FEARS STAFF OR PATIENT COULD BE HURT FROM WARMER LEAKING OUT. NO INJURY TO PATIENT OR STAFF FOR THIS INCIDENT.
Additional Manufacturer Narrative
THE COMPLAINT WAS FORWARDED TO THE MANUFACTURING FACILITY WHERE IT IS CURRENTLY STILL UNDER INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE THE RESULTS HAVE BEEN COMPLETED.
Additional Manufacturer Narrative
ONE SAMPLE WAS RECEIVED FOR EVALUATION. THE SAMPLE WAS ALREADY ACTIVATED; THEREFORE, THE ROOT CAUSE COULD NOT BE CONFIRMED. DEVICE HISTORY RECORD REVIEW WAS COMPLETED ON THE REPORTED LOT V3D237, AND THE LOT WAS MANUFACTURED AND RELEASED IN COMPLIANCE WITH ALL REQUIREMENTS. CARDINAL HEALTH HAS MADE A BUSINESS DECISION TO CLOSE THE SITE WHICH MANUFACTURES PRODUCT 11460-010T AND TRANSITION TO A THIRD-PARTY MANUFACTURER.