inforMED
InjuryQJH

SURFACER®

Received Jul 25, 2023 · Event occurred Jul 17, 2023

Report 1721504-2023-00065 · MDR key 17393786

Device

Generic name

Reverse Central Venous Recanalization System

Catalog number

600200/A

Lot number

BVT041923-01

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

Unknown

  • Perforation of Vessels
  • Perforation of Vessels

Narrative

Additional Manufacturer Narrative

THE SUSPECT DEVICE WAS NOT RETURNED FOR EVALUATION. THE COMPLAINT COULD NOT BE CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED. A SEARCH OF THE COMPLAINT DATABASE WAS PERFORMED AND NO SIMILAR COMPLAINTS FOR THIS LOT NUMBER WERE FOUND. THE DEVICE HISTORY RECORD WAS REVIEWED, AND NO EXCEPTION DOCUMENTS WERE FOUND.

Additional Manufacturer Narrative

THE SUSPECT DEVICE IS NOT EXPECTED TO RETURN FOR EVALUATION. A FOLLOW UP WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem

THE ACCOUNT ALLEGES THAT DURING A PROCEDURE, THE PATIENT EXPERIENCED AN ARTERIAL PERFORATION REQUIRING INTERVENTION AND ALSO SUFFERED A STROKE POST PROCEDURE DUE TO PHYSICIAN FAILURE TO PROPERLY ANTICOAGULANT THE PATIENT. THE PHYSICIAN STATES THAT THE DEVICE WAS SUCCESSFULLY ADVANCED FROM THE PATIENT'S RIGHT FEMORAL VEIN TO THE SITE OF THE OCCLUSION IN THE SVC. THE DEVICE WAS THEN ADVANCED THROUGH THE TARGETED OCCLUSION, AND THE TIP OF THE DEVICE WAS ADVANCED SUPERIOR TO THE RIGHT CLAVICLE. THE TIP OF THE DEVICE AND THE EXIT SITE TARGET WERE CORRECTLY ALIGNED UNDER FLUOROSCOPY AND THE NEEDLE WIRE WAS ADVANCED AND EXTERIORIZED WITH THE ASSISTANCE OF AN #11 BLADE AND A PAIR OF HEMOSTATS. THE PEEL AWAY SHEATH WAS THEN ADVANCED OVER THE WIRE TO THE PATIENT. HEMOSTATS WERE PLACED AT THE PROXIMAL END OF THE SHEATH CLAMPING THE WIRE IN PLACE. THE PHYSICIAN THEN LOCKED THE WIRE IN PLACE ON THE DEVICE AND PROCEEDED TO PULL THE SHEATH INSIDE THE PATIENT. THE PHYSICIAN THEN ADMINISTERED CONTRAST THROUGH THE SHEATH AND OBSERVED VASCULAR EXTRAVASATION DUE TO PERFORATION OF THE ARTERY. THE DEVICE WAS THEN REMOVED FROM THE PATIENT ENTIRELY AND THE INTERVENTIONAL CARDIOLOGY TEAM WAS URGENTLY CALLED TO ASSIST. THE TEAM SUCCESSFULLY DEPLOYED VASCULAR STENTS TO REPAIR THE ARTERIAL PERFORATION. THE PHYSICIAN BELIEVES HE INADVERTENTLY PERFORATED THE RIGHT INNOMINATE ARTERY WHILE MANIPULATING THE SURFACER DEVICE. HE ALSO BELIEVES HIS ACTIONS AND FAILURE TO PUSH HEPARIN DURING THE REST OF PROCEDURE MAY HAVE CAUSED THE PATIENT TO STROKE. THE EVENT WAS NOT A RESULT OF THE DEVICE MALFUNCTION. THE DEVICE PERFORMED AS INTENDED. THE PATIENT WAS TREATED WITH BENADRYL PRE-PROCEDURE DUE TO ALLERGIES TO VANCOMYCIN AND VARIOUS CONTRAST AGENTS.