ELECSYS IGE II
Received Jul 20, 2023 · Event occurred Jul 1, 2023
Report 1823260-2023-02366 · MDR key 17362987
Device
Generic name
Radioimmunoassay, Immunoglobulins (d,e)
Manufacturer
Roche DiagnosticsModel number
NACatalog number
04827031190
Lot number
67239400
Product problems
- Low Test Results
- Non Reproducible Results
- Low Test Results
- Non Reproducible Results
Patient
9 YR · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE CALIBRATION SIGNALS WERE WITHIN EXPECTATIONS. QUALITY CONTROLS PERFORMED ON 29-JUN-2023 AND ON 03-JUL-2023 WERE WITHIN RANGE. THE CUSTOMER DID NOT RUN CONTROLS ON THE DAY OF THE EVENT BASED ON THE PROVIDED INFORMATION. UPON REVIEW OF THE ALARM TRACE, NO RELEVANT ALARMS WERE OBSERVED. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
THE SERIAL NUMBER OF THE COBAS 6000 E601 MODULE IS (B)(6). THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ELECSYS IGE II ON A COBAS 6000 E601 MODULE. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN AN IGE VALUE OF 0.195 UI/ML ACCOMPANIED BY A DATA FLAG INDICATING CARRYOVER. THERE WERE ONLY 3 TESTS REMAINING IN THE REAGENT PACK DURING THIS MEASUREMENT. THE SAMPLE WAS REPEATED, RESULTING IN AN IGE VALUE OF 958.2 UI/ML. THIS MEASUREMENT WAS PERFORMED USING A NEW REAGENT PACK. AN ADDITIONAL IGE VALUE OF < 0.100 UI/ML WITH A DATA FLAG WAS PROVIDED. IT WAS ASKED, BUT IT IS NOT KNOWN IF THIS IS AN ADDITIONAL REPEAT VALUE FROM THE COMPLAINED SAMPLE OR IF IT IS A RESULT OF A DIFFERENT SAMPLE.