inforMED
MalfunctionILK

6257 XPEDITION POWERED CHAIR

Received Jul 14, 2023 · Event occurred Jul 3, 2023

Report 0001831750-2023-00684 · MDR key 17325309

Device

Generic name

Transport, Patient, Powered

Model number

6257

Catalog number

625700000000

Product problems

  • Difficult or Delayed Positioning
  • Difficult or Delayed Positioning

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE DEVICE GOT STUCK ON THE WAY UP A SLIGHTLY WINDING STAIRCASE. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Additional Manufacturer Narrative

AFTER INVESTIGATION, IT WAS DETERMINED THAT THE ISSUE OCCURRED DURING A DEMONSTRATION OF THE CHAIR. THE USER WAS TOO CLOSE TO THE RAILING ON A WINDING STAIRCASE, AND THE TRACKS CAUGHT PART OF THE RAILING. A VISUAL AND FUNCTIONAL INSPECTION WAS PERFORMED BY A STRYKER FIELD SERVICE TECHNICIAN. IT WAS FOUND THAT THE DIFFICULT TO MANEUVER CHAIR WAS DUE TO USER ERROR, RESULTING IN A BENT TRACK RAIL GUIDE CAGE. THIS ISSUE WAS RESOLVED FOR THE CUSTOMER BY REPLACING THE TRACK GUIDE.

Description of Event or Problem