inforMED
MalfunctionKRL

BUBBLE DETECTOR SENSOR

Received Jul 14, 2023 · Event occurred Jun 14, 2023

Report 9611109-2023-00330 · MDR key 17323000

Device

Generic name

Monitor And/or Control, Level Sensing, Cardiopulmonary Bypass

Model number

23-07-40

Product problems

  • Over-Sensing
  • Low Readings
  • Over-Sensing
  • Low Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

SEE INITIAL REPORT.

Additional Manufacturer Narrative

H10: THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE SENSOR REACTED EVEN THOUGH THERE WAS NO AIR BUBBLES. IN CASE OF BUBBLE SENSOR TRIGGERING FALSE BUBBLE ALARMS (OVER-SENSING ISSUE), THE PUMP IS STOPPED BY THE BUBBLE ALARM EVEN IF NO AIR IS PRESENT IN THE MONITORED LINE. IN SUCH SITUATIONS, THE ALARM CAN ALWAYS BE CLEARED/OVERRIDDEN BY USER, ONCE THE LINE IS VERIFIED TO BE FREE OF AIR, AND THE PERFUSION CAN BE EASILY RESTARTED. THUS, IT IS UNLIKELY THAT A FALSE BUBBLE ALARM WILL RESULT IN SERIOUS INJURY. BASED ON THE ABOVE RATIONALE, THE EVENT HAS BEEN RE-ASSESSED AS NOT REPORTABLE.

Additional Manufacturer Narrative

THERE WAS NO PATIENT INVOLVEMENT. LIVANOVA DEUTSCHLAND MANUFACTURES THE BUBBLE DETECTOR SENSOR. THE INCIDENT OCCURRED IN KURUME, JAPAN. THROUGH FOLLOW-UP COMMUNICATION WITH LIVANOVA FIELD SERVICE REPRESENTATIVE, A NEW BUBBLE DETECTOR SENSOR WILL BE SHIPPED TO THE CUSTOMER. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A BUBBLE SENSOR DETECTOR MALFUNCTIONED DURING PRIMING. NO ADDITIONAL INFORMATION IS AVAILABLE. IT CANNOT BE EXCLUDED THAT BUBBLE SENSOR DID NOT DETECT AIR BUBBLE. THERE WAS NO PATIENT INVOLVEMENT.

Remedial action

  • Other