BIOPOLY GREAT TOE HEMIARTHROPLASTY IMPLANT
Received Jul 11, 2023 · Event occurred Jun 16, 2023
Report 3008108789-2023-00005 · MDR key 17299738
Device
Generic name
Toe Prosthesis
Manufacturer
Biopoly LlcModel number
135-5118Catalog number
135-5118
Lot number
2133830
Product problems
- Off-Label Use
- Use of Device Problem
- Off-Label Use
- Use of Device Problem
Patient
37 YR · Male
- Osteolysis
- Osteolysis
Narrative
Additional Manufacturer Narrative
BIOPOLY RECEIVED THE EXPLANT FOR INSPECTION. THE EXPLANT WAS EXAMINED UNDER MAGNIFICATION AND APPEARED TO SHOW SIGNS OF ABRASIVE WEAR. THE REMAINING BIOPOLY MATERIAL WAS SECURELY ATTACHED TO THE TITANIUM STEM AND THE OSTEOSYNC MATERIAL WAS ALSO SECURELY ATTACHED. THE IMPLANT HAD BEEN SECURED IN THE BONE VIA PRESS FIT; HOWEVER, THERE WAS NO EVIDENCE THAT BONE CEMENT HAD BEEN USED. THE IMPLANT MANUFACTURING LOT MET ALL SPECIFICATIONS. ALTHOUGH IT IS NOT CONFIRMABLE, THE CAUSE APPEARS TO HAVE BEEN OVERSTUFFING OF THE JOINT ALONG WITH THE PHALANX BEING DEVOID OF CARTILAGE, A CONTRAINDICATION, LEADING TO EXCESSIVE AND ABRASIVE BIOPOLY WEAR AND LEADING TO OSTEOLYSIS. WEAR AND DAMAGE OF THE IMPLANT ARE POSSIBLE ADVERSE EFFECTS LISTED IN THE INSTRUCTIONS FOR USE (IFU) DOCUMENT AND CONSIDERED RISKS IN THE RISK MANAGEMENT SYSTEM FOR THIS IMPLANT. MANY POSSIBLE FACTORS COULD LEAD TO DAMAGE TO THE IMPLANT DURING IMPLANTATION, INCLUDING FAILURE TO FOLLOW INSTRUCTIONS FOR USE, USE OF IMPLANT AGAINST NON-CARTILAGE SURFACE, OR IMPROPER PATIENT SELECTION (AGE, BONE QUALITY, CARTILAGE HEALTH). IF ADDITIONAL INFORMATION IS RECEIVED, INFORMATION WILL BE REVIEWED FOR REPORTABILITY AND SUBMITTED FOR FOLLOW-UP AS APPROPRIATE.
Description of Event or Problem
BIOPOLY LEARNED FROM PHYSICIAN ABOUT OSTEOLYSIS/BONE RESORPTION IN ONE OF HIS PATIENTS, AND HE REMOVED THE BIOPOLY IMPLANT.