Narrative

Description of Event or Problem

FACE MASKS INSIDE THE KIT ARE FLAT(UNDERINFLATED).

Additional Manufacturer Narrative

THE MASK DEFLATING AND NOT PROPERLY SEALING TO PATIENT WOULD CAUSE A DISRUPTION IN RESUSCITATION EFFORTS AND PUT THEM AT RISK. SUMMARY: COMPLAINT CONFIRMED VIA RETURNED DEVICES. THE MASKS ARE ABLE TO BE INFLATED AND USED PROPERLY. THE CUSTOMER WAS NOTIFIED OF POTENTIAL CAUSES FOR DEFLATED MASKS. RA: INSTRUCTIONS FOR MASK USE IN COMBINATION WITH RMA-20020 ARE SUFFICIENT RISK ANALYSIS FOR THE REPORTED PRODUCT COMPLAINT.

Description of Event or Problem

FACE MASKS INSIDE THE KIT ARE FLAT (UNDERINFLATED).

Additional Manufacturer Narrative

THE MASK DEFLATING AND NOT PROPERLY SEALING TO PATIENT WOULD CAUSE A DISRUPTION IN RESUSCITATION EFFORTS AND PUT THEM AT RISK.