inforMED
MalfunctionOSN

MAXFRAME AUTOSTRUT(TM)HEXAPOD STRUT-MED

Received Jul 6, 2023 · Event occurred Jan 1, 2023

Report 3015781803-2023-00020 · MDR key 17265567

Device

Generic name

Software For Diagnosis/treatment

Model number

1100010-01

Catalog number

1100010-01

Product problems

  • Break
  • Break

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: THE PHOTO WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE FROM THE PHOTO. VISUAL ANALYSIS OF THE PHOTO REVEALED THAT THE MAXFRAME AUTOSTRUT(TM)HEXAPOD STRUT-MED HAD NO ISSUE. THE DEVICE DOES NOT PRESENT ANY SIGNS OF BREAKAGE OR DEFECT, HENCE THE COMPLAINT CONDITION CANNOT BE CONFIRMED. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED FOR THE MAXFRAME AUTOSTRUT(TM)HEXAPOD STRUT-MED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED ON AN UNKNOWN DATE, THE PATIENT CAME INTO CLINIC WITH THREE BROKEN STRUTS. THE PATIENT CLAIMED THAT HE WAS SITTING BY SPACE HEATER WHEN THREE OF HIS MEDIUM AUTOSTRUTS CRACKED AT THE BASE OF BOLT CONNECTION. HE IS NOT BEARING ANY WEIGHT AND THIS IS THE SECOND OCCURRENCE OF THIS EVENT. PATIENT CLAIMED TO NOT BE TIGHTENING THE STRUTS HIMSELF. THERE WAS NO PATIENT OUTCOME OR CONSEQUENCES. THIS REPORT IS FOR ONE (1) MAXFRAME AUTOSTRUT(TM)HEXAPOD STRUT-MED THIS IS REPORT 2 OF 3 FOR COMPLAINT (B)(4).

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D2B: ADDITIONAL PRODUCT CODE: KTT. D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. E3: REPORTER IS A J&J SALES REPRESENTATIVE. H3, H4, H6: WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.