NOVAPLUS INFANT HEEL WARMER SQUEEZE
Received Jun 30, 2023 · Event occurred Jun 1, 2023
Report 1423537-2023-00957 · MDR key 17243373
Device
Generic name
Infant Heel Warmer (chemical Heat Pack)
Manufacturer
Mo016 Mo-moberlyModel number
V11460-010Catalog number
V11460-010
Lot number
UNKNOWN
Product problems
- Burst Container or Vessel
- Burst Container or Vessel
Patient
25 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS COMPLAINT WAS FORWARDED TO THE MANUFACTURING FACILITY WHERE IT IS CURRENTLY STILL UNDER INVESTIGATION. A FOLLOW UP REPORT WILL BE FILED ONCE THE RESULTS HAVE BEEN COMPLETED.
Description of Event or Problem
IT HAS BEEN REPORTED THAT WHEN A NURSE ATTEMPTED TO ACTIVATE A HOT PACK USING NORMAL PRESSURE THE PACK CAME APART. THE BROKEN SIDE WAS FACING UPWARDS AND AS IT BROKE IT BURST FORTH ITS CONTENTS SPRAYING ONTO THE NURSE AND ACROSS TWO OCCUPIED BASSINETS. SOME OF THE CONTENTS SPLASHED ONTO THE NURSES HANDS AND ON HER CHEST. THE CONTENTS WERE IMMEDIATELY CLEANED UP WITHOUT ANY INTERACTION. NURSE IS UNAFFECTED BY THE EVENT.
Additional Manufacturer Narrative
CARDINAL HEALTH HAS RECEIVED AN INCREASE IN BURST/LEAKING COMPLAINTS REGARDING THE NOVALPLUS INFANT HEEL WARMER WITH TAPE. CUSTOMER COMPLAINTS HAVE NOTED RUPTURING AND BURSTING OF THE DEVICE WHEN ACTIVATION IS ATTEMPTED, MAKING THE DEVICE UNUSABLE AND IN MANY CASES CAUSING THE CONTENTS INSIDE THE POUCH TO LEAK AND CONTACT CLINICIANS, PATIENTS, AND OTHER INDIVIDUALS OR SURFACES IN PROXIMITY. SAMPLES RETURNED TO CARDINAL HEALTH FROM CUSTOMERS HAVE BEEN EVIDENCED TO HAVE AN INCOMPLETE TOP SEAL, RESULTING IN THE MALFUNCTION. THERE HAVE BEEN NO REPORTABLE ADVERSE EVENTS/INJURIES ASSOCIATED WITH V11460-010, LOT V2S056. CARDINAL HEALTH HAS HALTED DISTRIBUTION OF THE NOVALPLUS INFANT HEEL WARMER W/TAPE FOR LOT NUMBER V2S056 AND INITIATED A VOLUNTARY RECALL ON JUNE 16, 2023.
Remedial action
- Recall