inforMED
MalfunctionGAS

ETHIBOND EXTRA & EXCEL POLYESTER SUTURE

Received Jun 7, 2010

Report 2210968-2010-00606 · MDR key 1723768

Device

Generic name

Suture, Nonabsorbable

Manufacturer

Ethicon, Inc.

Model number

NA

Catalog number

EH7579H

Lot number

CA2JBWM0

Product problems

  • Break

Patient

UNK

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT A PATIENT UNDERWENT AN EAR, NOSE AND THROAT SEPTUM DEFECT PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE NEEDLE BROKE DURING USE ON THE PATIENT. THERE WAS NO ADVERSE PATIENT CONSEQUENCE.

Additional Manufacturer Narrative

(B)(4): RESULTS: REPRESENTATIVE SAMPLES OF THE PRODUCT WERE RETURNED FOR EVALUATION. THE NEEDLES WERE INSPECTED AND TESTED FOR STRENGTH. THERE WERE NO SIGNS OF MANUFACTURING OR MATERIAL DEFECTS FOUND. CONCLUSION: NO DEVICE FAILURE DETECTED AND THE PRODUCT IS WITHIN SPECIFICATION.