inforMED
InjuryFBO

UROLOGY INNER SHEATH

Received Jun 27, 2023 · Event occurred May 25, 2023

Report 2020550-2023-00135 · MDR key 17213291

Device

Generic name

Inner Resectoscope Sheath

Model number

27050CA

Catalog number

27050CA

Lot number

TQ06

Product problems

  • Break
  • Detachment of Device or Device Component
  • Break
  • Detachment of Device or Device Component

Patient

69 YR · Male

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION WITH A 27050CA INNER SHEATH FOR 27050SC. ACCORDING TO THE INFORMATION RECEIVED, THE CERAMIC BEAK BROKE INSIDE PATIENTS BLADDER DURING A TURP PROCEDURE. THE PIECE/PIECES WERE ABLE TO BE REMOVED WITH NO DIRECT HARM TO PATIENT, HOWEVER; AN ADDITIONAL 30 MINUTES OF ANESTHESIA WAS ADDED TO THE PROCEDURE. IT WAS NOTED THE POSSIBILITY OF A 3RD PARTY REPAIR WAS OBSERVED BY SE.

Additional Manufacturer Narrative

MANUFACTURING SITE EVALUATION: THE CERAMIC DISTAL TIP IS BROKEN/FRACTURED WITH REMAINING ATTACHED BEAK HAVING SHARP, JAGGED EDGE. LOT CODE IS TQ 06((06/2019). SHOULD RELEVANT ADDITIONAL INFORMATION/INVESTIGATION RESULTS BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UNSOLICITED. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID (B)(4).