PFR95 PART FILT RESP
Received Jun 20, 2023 · Event occurred Sep 25, 2020
Report 9616096-2023-00064 · MDR key 17167922
Device
Generic name
Icp Facial Protection
Manufacturer
O&m Halyard, Inc.Model number
46727Catalog number
991046727
Lot number
AM0070821
Product problems
- Break
- Break
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS PRODUCT WAS NOT RETURNED FOR EVALUATION. THE DEVICE HISTORY RECORDS (DHR) EVALUATION WAS PERFORMED FOR PRODUCT CODE 46727 IDENTIFIED IN THE COMPLAINT. ACCORDING TO THE DOCUMENTED RECORDS, THE PRODUCT WAS MANUFACTURED ACCORDING TO THE APPROVED MANUFACTURING PROCEDURES AND SPECIFICATIONS. THE PRODUCT WAS RELEASED BY QUALITY ASSURANCE. A QUALITY NONCONFORMANCE WAS NOT REPORTED AT THE TIME OF PRODUCTION. NO ROOT CAUSE WAS IDENTIFIED. MANUFACTURING CONDUCTS VISUAL AND FUNCTIONAL INSPECTIONS THROUGHOUT PRODUCTION ASIDE FROM THE QA INSPECTION. NOTIFICATION WAS SENT TO MANUFACTURING LEADERS FOR AWARENESS. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE O&M HALYARD, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN O&M HALYARD, INC. PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY.
Description of Event or Problem
WE HAVE EXPERIENCED BOTH REGULAR AND SMALL SIZES BREAKING IN THE STRAP AREA CONNECTED TO THE MASK. INCIDENT WAS REPORTED TO HAVE OCCURRED DURING PRODUCT USE. THE FOLLOWING MEDICAL DEVICE REPORT IS BEING FILED RETROACTIVE TO PRIOR DECISION THAT THIS MALFUNCTION WOULD BE UNLIKELY TO RESULT IN A SERIOUS INJURY OR ILLNESS. A REASSESSMENT OF THIS INCIDENT HAS SINCE FOUND THAT SHOULD THIS MALFUNCTION RECUR IN AN ISOLATION SETTING OR DURING DIRECT PATIENT CARE, THE END USER MAY CONTRACT COVID VIRUS AND THUS BE SUSCEPTIBLE TO A SERIOUS ILLNESS. FOR THIS REASON THIS INCIDENT IS BEING REPORTED AS A MALFUNCTION AT THIS TIME.