BD BBL¿ BRAIN HEART INFUSION

Received Jun 16, 2023 · Event occurred May 30, 2023

Report 1119779-2023-00675 · MDR key 17148459

Device

Generic name

Culture Media, General Nutrient Broth

Model number

220837

Catalog number

220837

Lot number

2279210

Product problems

  • Contamination of Device Ingredient or Reagent
  • Contamination of Device Ingredient or Reagent

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT BD BBL¿ BRAIN HEART INFUSION MEDIA WAS CONTAMINATED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VISIBLE CONTAMINATION. TUBES SHOW TURBIDITY WHEN DID THE INCIDENT OCCUR?: BEFORE USE UPDATE- 12/JUN/2023: CUSTOMER REFER TO HAVE SEEN BACILLI ON SLIDE ALTHOUGH NOTHING GREW ON PLATES.

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: MATERIAL 220837 ARE MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER, AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE TUBES ARE FILLED, CAPPED, TORQUED, AND THEN LABELED BY MACHINE PER STANDARD OPERATING PROCEDURE (SOP). THE TUBES ARE TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, TUBES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE BATCH HISTORY RECORD REVIEW FOR BATCH 2279210 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. FORMULATION AND FILLING PROCESSES WERE WITHIN SPECIFICATIONS. IN PROCESS CHECKS WERE PERFORMED AT THE DESIGNATED INTERVALS. THOSE CHECKS CONFIRMED THAT THE CAPS WERE TIGHTENED TO THE VALIDATED SPECIFICATIONS PER INTERNAL PROCEDURE. QC INSPECTION AND TESTING WERE SATISFACTORY AT TIME OF RELEASE. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE REVIEW OF ITS COLOR AND CLARITY. SAMPLES SUBMITTED ARE EXAMINED TO ENSURE THAT THEY CONFORM TO TYPICAL LEVELS. THE APPEARANCE OF THIS BATCH WAS SATISFACTORY PER INTERNAL PROCEDURES. DIRECT STAINING TECHNIQUES ARE NOT PART OF QC RELEASE TESTING FOR THIS PRODUCT. ADDITIONALLY, AS PART OF THE RELEASE CRITERIA FOR THIS PRODUCT, THE BHR IS REVIEWED TO CONFIRM THE FOLLOWING: --THE TOTAL ELAPSED TIME BETWEEN END OF FORMULATION AND START OF THE AUTOCLAVE CYCLE WAS WITHIN THE SPECIFIED LIMITS. --ALL AUTOCLAVE PARAMETERS CONFORMED TO THE VALIDATED CYCLE PARAMETERS FOR THIS PRODUCT. --THE MINIMUM F0 FOR THIS PRODUCT WAS MET. THE COMPLAINT HISTORY WAS REVIEWED, AND FOUR OTHER COMPLAINTS HAVE BEEN TAKEN ON THIS BATCH. RETENTION SAMPLES FROM BATCH 2279210 (10 TUBES) WERE AVAILABLE FOR INSPECTION. NO MEDIA DEFECTS WERE OBSERVED IN 10/10 RETENTION SAMPLES. ALL RETENTIONS TUBES HAD THE EXPECTED APPEARANCE FOR THIS PRODUCT OF CLEAR TO TRACE HAZY, LIGHT TO MEDIUM DARK YELLOW TAN AS DESCRIBED IN THE CERTIFICATE OF ANALYSIS. FOR INVESTIGATION, TWO RETENTION TUBES WENT INTO INCUBATION. ONE RETENTION TUBE WAS PLACED INTO THE 20¿25-DEGREE CELSIUS INCUBATOR AND ONE RETENTION TUBE WAS PLACED IN THE 33¿37-DEGREE CELSIUS INCUBATOR. AT THE SEVENTH DAY OF INCUBATION THERE WERE NO SIGNS OF GROWTH OR TURBIDITY OR CHANGE IN MEDIA COLOR AND CLARITY. AFTER INCUBATION THE BROTH APPEARANCE REMAINED CLEAR TO TRACE HAZY, LIGHT TO MEDIUM DARK YELLOW AS DESCRIBED IN THE CERTIFICATE OF ANALYSIS. ONE PHOTO WAS RECEIVED TO ASSIST WITH THE INVESTIGATION: THE PHOTO SHOWS ELEVEN TUBES FROM BATCH 2279210 IN A TUBE RACK. THE MEDIA APPEARS CLEAR TO TRACE HAZY, LIGHT TO MEDIUM DARK YELLOW TAN AS DESCRIBE IN THE CERTIFICATE OF ANALYSIS. NO RETURNS WERE RECEIVED TO ASSIST WITH THE INVESTIGATION. THIS COMPLAINT CANNOT BE CONFIRMED BASED ON THE EVIDENCE PROVIDED BY THE PHOTO RECEIVED. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION/APPEARANCE/NON-VIABLE DEFECTS. CAUTION SHOULD BE EXERCISED IN REPORTING DIRECT GRAM STAIN AND/OR OTHER DIRECT MICROBIOLOGICAL STAIN RESULTS ON TISSUE SPECIMENS PROCESSED WITH THIS MEDIUM DUE TO THE POSSIBLE PRESENCE OF NONVIABLE ORGANISMS IN THE CULTURE MEDIUM. CULTURE MEDIA SOMETIMES CONTAIN DEAD ORGANISMS DERIVED FROM MEDIUM CONSTITUENTS, WHICH MAY BE VISIBLE IN SMEARS OF CULTURE MEDIA. OTHER SOURCES OF DEAD ORGANISMS VISIBLE UPON GRAM STAINING INCLUDE STAINING REAGENTS, IMMERSION OIL, GLASS SLIDES, AND THE SPECIMENS USED FOR INOCULATION. IF THERE IS UNCERTAINTY ABOUT THE VALIDITY OF THE GRAM STAIN, THE CULTURE SHOULD BE REINCUBATED FOR ANOTHER HOUR OR TWO AND THE TEST REPEATED BEFORE A REPORT IS GIVEN. BD WILL CONTINUE TO TREND COMPLAINTS FOR APPEARANCE AND NON-VIABLES. THIS COMPLAINT CANNOT BE CONFIRMED.

Additional Manufacturer Narrative

E1. INITIAL REPORTER PREFIX: (B)(6). INITIAL REPORTER FACILITY NAME: (B)(6). H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT WAS REPORTED THAT BD BBL¿ BRAIN HEART INFUSION MEDIA WAS CONTAMINATED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VISIBLE CONTAMINATION. TUBES SHOW TURBIDITY. WHEN DID THE INCIDENT OCCUR?: BEFORE USE. UPDATE- 12/JUN/2023: CUSTOMER REFER TO HAVE SEEN BACILLI ON SLIDE ALTHOUGH NOTHING GREW ON PLATES.