V-PRO MAX 2 STERILIZER

Received Jun 16, 2023 · Event occurred May 19, 2023

Report 3005899764-2023-00029 · MDR key 17144767

Device

Generic name

Sterilizer

Model number

MAX 2

Catalog number

VP50002101

Product problems

  • Use of Device Problem
  • Use of Device Problem

Patient

Unknown

  • Burn(s)
  • Burn(s)

Narrative

Additional Manufacturer Narrative

THE ACCOUNT MANAGER OFFERED IN-SERVICE TRAINING ON THE PROPER USE OF HANDLING INSTRUMENT PACKS, SPECIFICALLY ENSURING ALL INSTRUMENTS ARE PROPERLY DRIED PRIOR TO PROCESSING. THE CUSTOMER DECLINED IN-SERVICE TRAINING. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A BURN TO THEIR HAND WHILE HANDLING AN INSTRUMENT PACK THAT WAS STERILIZED IN THEIR V-PRO MAX 2 STERILIZER. THE EMPLOYEE SOUGHT AND RECEIVED MEDICAL TREATMENT.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ON SITE TO INSPECT THE V-PRO MAX 2 STERILIZER AND FOUND IT TO BE OPERATING PROPERLY. NO ISSUES WITH THE FUNCTION OR OPERATION OF THE STERILIZER WERE IDENTIFIED. THE TECHNICIAN WAS INFORMED THAT THE INSTRUMENT PACK HAD BEEN PROCESSED IN THE STERILIZER AND THEN PLACED ON A DRYING RACK IN THE SURGICAL ROOM. PRIOR TO THE START OF A PROCEDURE, A NURSE IDENTIFIED LIQUID ON THE OUTSIDE OF THE INSTRUMENT PACK, THE LIQUID THEN MADE CONTACT WITH THE NURSE'S HAND RESULTING IN THE REPORTED EVENT. THE V-PRO MAX 2 STERILIZER OPERATOR MANUAL (1-2) STATES, "DANGER - CHEMICAL INJURY HAZARD: ANY VISIBLE LIQUIDS IN THE CHAMBER OR ON THE LOAD MUST BE TREATED AS CONCENTRATED HYDROGEN PEROXIDE. OBSERVE ALL HYDROGEN PEROXIDE HANDLING PRECAUTIONS." USER FACILITY PERSONNEL SHOULD ENSURE ALL INSTRUMENTS ARE PROPERLY DRY PRIOR TO PLACEMENT IN THE V-PRO MAX 2 STERILIZER. THE V-PRO MAX 2 STERILIZER OPERATOR MANUAL, (A-1) STATES, "DRY ALL ITEMS THOROUGHLY. ENSURE ALL MOISTURE IS REMOVED FROM ALL INTERNAL PARTS (INCLUDING LUMENS). IF NOT, RESIDUAL HYDROGEN PEROXIDE MAY REMAIN AT CYCLE COMPLETION AND/OR A CYCLE ABORT OCCURS. ONLY DRY ITEMS ARE TO BE PLACED IN STERILIZATION UNIT." THE ACCOUNT MANAGER WILL PROVIDE IN-SERVICE TRAINING TO FACILITY PERSONNEL ON THE PROPER USE OF HANDLING OF INSTRUMENT PACKS, SPECIFICALLY ENSURING ALL INSTRUMENTS ARE PROPERLY DRIED PRIOR TO PROCESSING. NO ADDITIONAL ISSUES HAVE BEEN REPORTED