STERRAD® 100NX STERILIZER 2-DR
Received Jun 14, 2023 · Event occurred Oct 25, 2022
Report 2084725-2023-00138 · MDR key 17132010
Device
Generic name
Sterrad® Equipment
Manufacturer
Advanced Sterilization ProductsModel number
10104-002Catalog number
10104-002
Lot number
NA
Product problems
- Environmental Particulates
- Environmental Particulates
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE FIELD SERVICE ENGINEER IDENTIFIED THAT THE OIL MIST FILTER SCREW WAS LOOSE. HE ADJUSTED/TIGHTENED THE OIL MIST FILTER SCREW TO RESOLVE THE MIST/HAZE ISSUE AND CONFIRMED THAT THE UNIT MET SPECIFICATIONS. THE SYSTEM WAS RETURNED TO SERVICE. H3: ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD (DHR), TRENDING ANALYSIS OF THE MIST/HAZE ISSUE, AND SYSTEM RISK ANALYSIS (SRA). THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ISSUES RELATING TO THE FAILURE MODE WERE NOTED. THE INVOLVED UNIT MET MANUFACTURER SPECIFICATIONS AT THE TIME OF RELEASE. TRENDING ANALYSIS OF THE MIST/HAZE ISSUE FOR THE STERRAD® 100NX UNIT WAS REVIEWED FOR THE PRIOR SIX MONTHS FROM THE EVENT DATE, AND NO SIGNIFICANT TREND WAS OBSERVED. REVIEW OF RISK DOCUMENTATION SHOWS THE RISK FOR EXPOSURE TO TOXIC OR CORROSIVE MATERIAL TO BE "LOW." NO PARTS WERE REPLACED IN RESOLVING THIS ISSUE; THEREFORE, NO PARTS WERE AVAILABLE FOR FURTHER ANALYSIS. THE ASSIGNABLE CAUSE OF THE MIST/HAZE IS POTENTIALLY DUE TO THE OIL MIST FILTER SCREW. THE FIELD SERVICE ENGINEER ADJUSTED/TIGHTENED THE PART AND CONFIRMED THE STERRAD® 100NX MET FUNCTIONAL SPECIFICATIONS AFTER SERVICE. THE ISSUE WAS RESOLVED AT THE CUSTOMER FACILITY. ASP WILL CONTINUE TO TRACK AND TREND THIS ISSUE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ADVANCED STERILIZATION PRODUCTS, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ADVANCED STERILIZATION PRODUCTS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ASP COMPLAINT REF #: (B)(4).
Description of Event or Problem
WHILE AT CUSTOMER SITE FOR A SEPARATE ISSUE, THE FIELD SERVICE ENGINEER REPORTED AN EVENT OF A ¿SMOKE¿ OR MIST/HAZE EMITTING FROM THE STERRAD® 100NX STERILIZER. THERE WAS NO REPORT OF ANY INJURIES OR HUMAN REACTIONS. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A PREVIOUS SERIOUS INJURY.