PFR95 PART FILT RESP
Received Jun 6, 2023 · Event occurred Oct 21, 2020
Report 9616096-2023-00030 · MDR key 17076223
Device
Generic name
Icp Facial Protection Products
Manufacturer
O&m Halyard, Inc.Model number
46727Catalog number
991046727
Lot number
AJ613116D
Product problems
- Material Integrity Problem
- Material Integrity Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
INCIDENT ONE: "SEVERAL DIFFERENT EPISODES OF THE STRAPS BREAKING ON THIS MASK BOTH DURING APPLICATIONS AS WELL AS MID STERILE PROCEDURE. TODAY WHILE SCRUBBED INTO A HEART CATH THE TOP STRAP BROKE ON THE MASK. DR.(B)(6) WOULD LIKE TO ADD THAT THIS ALSO HAPPENED DURING A PROCEDURE ON A CODE HEART WITH UNKNOWN COVID STATUS POTENTIALLY EXPOSING THE PHYSICIAN TO COVID. THIS IS OCCURRING DESPITE PRE-APPLICATION STRETCHING AS WELL NO PRE-APPLICATION STRETCHING." THE FOLLOWING MEDICAL DEVICE REPORT IS BEING FILED RETROACTIVE TO PRIOR DECISION THAT THIS MALFUNCTION WOULD BE UNLIKELY TO RESULT IN A SERIOUS INJURY OR ILLNESS. A REASSESSMENT OF THIS INCIDENT HAS SINCE FOUND THAT SHOULD THIS MALFUNCTION RECUR IN AN ISOLATION SETTING OR DURING DIRECT PATIENT CARE, THE END USER MAY CONTRACT COVID VIRUS AND THUS BE SUSCEPTIBLE TO A SERIOUS ILLNESS. FOR THIS REASON THIS INCIDENT IS BEING REPORTED AS A MALFUNCTION AT THIS TIME. THE REPORTER HAS INDICATED THAT NO INJURY OR ILLNESS RESULTED.
Additional Manufacturer Narrative
FOUR (4) KIMBERLY CLARK BRANDED SAMPLES WERE RECEIVED. ONE KIMBERLY CLARK PRODUCT BOX WAS RECEIVED. THE SAMPLES WERE EVALUATED UNDER AMBIENT LIGHT CONDITIONS. THE SAMPLES WERE OPENED AND INSPECTED. EACH OF THE SAMPLES ELASTIC HEAD BANDS APPEAR STUCK TOGETHER. THE ELASTICS FEEL STIFF AND BRITTLE. THE ELASTIC OF ONE SAMPLE WAS TUGGED AS IF FOR DONNING. THE ELASTIC BROKE IN THE CENTER OF THE BANDS, DETACHING A SECTION OF THE ELASTIC. THIS PRODUCT IS OVER 14 YEARS OLD FROM THE LOT PROVIDED AJ613116D. CODE 46727 MENTIONED IN COMPLAINT HAS A SHELF LIFE OF 5 YEARS. THIS PRODUCT HAS EXPIRED. NOTIFICATION WAS SENT TO MANUFACTURING LEADERS FOR AWARENESS. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE O&M HALYARD, INC. COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN O&M HALYARD, INC. PRODUCT IS DEFECTIVE OR HAS CAUSED SERIOUS INJURY.