inforMED
MalfunctionMPO

NOVAPLUS INFANT HEEL WARMER SQUEEZE

Received Jun 5, 2023 · Event occurred May 5, 2023

Report 1423537-2023-00945 · MDR key 17063388

Device

Generic name

Infant Heel Warmer (chemical Heat Pack)

Manufacturer

Mo016 Mo-moberly

Model number

V11460-010

Catalog number

V11460-010

Lot number

V28056 E

Product problems

  • Burst Container or Vessel
  • Burst Container or Vessel

Patient

30 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

DURING A VIRTUAL IN-SERVICE WITH THE TEAM AT (B)(6), THE CUSTOMER IS REPORTING THAT THEY HAD INFANT HEEL WARMERS BURST WHILE ATTEMPTING TO ACTIVATE THEM. NO ONE WAS INJURED AS THESE WERE CONTROLLED TESTS.

Additional Manufacturer Narrative

THIS COMPLAINT WAS FORWARDED TO THE MANUFACTURING FACILITY WHERE IT IS CURRENTLY STILL UNDER INVESTIGATION. A FOLLOW UP REPORT WILL BE FILED ONCE THE RESULTS HAVE BEEN COMPLETED.

Additional Manufacturer Narrative

CARDINAL HEALTH HAS RECEIVED AN INCREASE IN BURST/LEAKING COMPLAINTS REGARDING THE NOVALPLUS INFANT HEEL WARMER WITH TAPE. CUSTOMER COMPLAINTS HAVE NOTED RUPTURING AND BURSTING OF THE DEVICE WHEN ACTIVATION IS ATTEMPTED, MAKING THE DEVICE UNUSABLE AND IN MANY CASES CAUSING THE CONTENTS INSIDE THE POUCH TO LEAK AND CONTACT CLINICIANS, PATIENTS, AND OTHER INDIVIDUALS OR SURFACES IN PROXIMITY. SAMPLES RETURNED TO CARDINAL HEALTH FROM CUSTOMERS HAVE BEEN EVIDENCED TO HAVE AN INCOMPLETE TOP SEAL, RESULTING IN THE MALFUNCTION. THERE HAVE BEEN NO REPORTABLE ADVERSE EVENTS/INJURIES ASSOCIATED WITH V11460-010, LOT V2S056. CARDINAL HEALTH HAS HALTED DISTRIBUTION OF THE NOVALPLUS INFANT HEEL WARMER W/TAPE FOR LOT NUMBER V2S056 AND INITIATED A VOLUNTARY RECALL ON (B)(6) 2023.

Remedial action

  • Recall