inforMED
MalfunctionKZF

LUCEA 40

Received Jun 1, 2023 · Event occurred May 26, 2023

Report 9710055-2023-00413 · MDR key 17036657

Device

Generic name

Device, Medical Examination, Ac Powered

Manufacturer

Maquet Sas

Model number

ARD568601998

Catalog number

ARD568601998

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

MANUFACTURER'S REFERENCE NUMBER (B)(4).

Additional Manufacturer Narrative

THE CORRECTION OF D4 VERSION OF MODEL AND D4 CATALOG # DEEMS REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS D4 VERSION OF MODEL AND D4 CATALOG # ARDLCA209000C. CORRECTED D4 VERSION OF MODEL AND D4 CATALOG # ARD568601998. GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EXAMINATION LIGHTS - LUCEA 40. IT WAS STATED THE CAP WAS MISSING FROM THE SPRING ARM. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF DURING EXAMINATION PROCEDURE MAY CAUSE CONTAMINATION. MISSING PART CAP SPRING ARM BLUE 30 NEW RAL9016 (ARD569010102) WAS REPLACED AND DEVICE BACK TO USAGE. BASED ON THE INFORMATION COLLECTED, IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, THE EXAMINATION LIGHT DID NOT MEET ITS SPECIFICATION, SINCE A MISSING CAP COVER COULD BE CONSIDERED AS TECHNICAL DEFICIENCY, AND IN THIS WAY THE DEVICE CONTRIBUTED TO THE EVENT. THE DEVICE WAS NOT BEING USED FOR PATIENT TREATMENT UPON THE EVENT OCCURRENCE. ACCORDING TO THE INFORMATION GATHERED, THE ISSUE WAS DISCOVERED BY GETINGE TECHNICIAN DURING PREVENTIVE MAINTENANCE. WHEN REVIEWING REPORTABLE EVENTS FOR THIS TYPE OF ISSUE WE WERE ABLE TO ESTABLISH THAT THE RECEIVED INCIDENTS ARE OCCURRING AT A VERY LOW RATIO. WE HAVE BEEN ABLE TO CONFIRM THAT THE INVESTIGATED ISSUE HAS NEVER LED TO SERIOUS INJURY OR WORSE, TO OUR KNOWLEDGE. AS STATED BY SUBJECT MATTER EXPERT AT MAQUET SAS, THE MOST PROBABLE ROOT CAUSE OF THE BREAK OF THIS CAP IS REPEATED AND VIOLENT SHOCKS DURING THE USE OF THE DEVICE. ANOTHER PROBABLE ROOT CAUSE IS THAT THE CAP HAS BEEN FORGOTTEN OR DETERIORATED AFTER A READJUSTMENT OF THE SPRING ARM DURING THE MAINTENANCE OF MEDICAL DEVICE. THE YEARLY PREVENTIVE MAINTENANCE PROGRAM DOCUMENTED IN THE TECHNICAL MANUAL FOR LUCEA 10/40 (0170201 1K) ON PAGE 10 MENTIONS TO CHECK THE FIXING OF ALL CAPS. THE CAP MUST BE REINSTALLED DURING INSTALLATION OR AFTER THE MAINTENANCE PROCEDURE. MAQUET SAS STRONGLY ADVISES TO CHECK SIMILAR DEVICES IN THE HOSPITAL IN ORDER TO CHECK THE PRESENCE OF ALL SPRING ARMS CAPS. IF A MISSING CAP IS NOTICED, A NEW ONE SHOULD BE ORDERED AS SPARE PARTS. (BLUE 30 / LUCEA 40-50 : ARD569010102) WE BELIEVE THE RELATED DEVICES ARE PERFORMING CORRECTLY IN THE MARKET. WE ALSO BELIEVE THAT IF THE MANUFACTURER RECOMMENDATION HAD BEEN FOLLOWED THE INCIDENT COULD HAVE BEEN AVOIDED. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. DEVICE NOT RETURNED TO MANUFACTURER.

Description of Event or Problem

ON (B)(6) 2023 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF EXAMINATION LIGHTS - LUCEA 40. IT WAS STATED THE CAP WAS MISSING FROM THE SPRING ARM. WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY PARTS FALLING OFF DURING EXAMINATION PROCEDURE MAY CAUSE CONTAMINATION.