SMALL PATCH DF-1 SUTURE UNI 67CM
Received May 21, 2010 · Event occurred Apr 7, 2010
Report 2124215-2010-10560 · MDR key 1703480
Device
Generic name
Implantable Lead
Manufacturer
Cpi - Del CaribeModel number
0067Catalog number
640067-107
Lot number
NA
Product problems
- High impedance
Patient
45 YR
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND CHRONIC EPICARDIAL PATCH DEFIBRILLATION LEADS WERE MEASURING A SHOCK IMPEDANCE OF GREATER THAN 125 OHMS. A TRIAD SHOCKING VECTOR DID NOT CONVERT THE PATIENT AT 21 J, AND THE PHYSICIAN REPORTEDLY BELIEVES THIS WAS DUE TO A DEVICE FAILURE. THE SHOCK VECTOR WAS REPROGRAMMED COIL TO COIL, RESULTING IN SUCCESSFUL CONVERSION AT 21 J. THE BOSTON SCIENTIFIC TECHNICAL SERVICES DEPARTMENT DISCUSSED THE POSSIBILITY OF DEFIBRILLATION LEAD PINS BEING REVERSED IN THE HEADER, A LEAD FRACTURE, OR A CONNECTION ISSUE. TECHNICAL SERVICES RECOMMENDED A 41 J AND 1.1 J COMMANDED SHOCK TO MEASURE LEAD IMPEDANCES AND TO VERIFY THERAPY DELIVERY. THE FOLLOWING CLINIC BROUGHT THE PATIENT IN FOR TROUBLESHOOTING AND DEFIBRILLATION THRESHOLD (DFT) TESTING. A 31 J SHOCK WAS DELIVERED, WHICH SUCCESSFULLY CONVERTED THE PATIENT. SHOCK IMPEDANCE MEASUREMENTS FOR THE SHOCK WERE WITHIN NORMAL RANGE, BUT RETURNED TO >125 OHMS, APPROXIMATELY 5 MINUTES LATER. AT THIS TIME, THE PHYSICIAN HAS NOT ELECTED TO PERFORM ANY ADDITIONAL REPROGRAMMING OR SURGICAL INTERVENTION. ADDITIONAL INFORMATION WAS PROVIDED THAT ADDITIONAL OUT OF RANGE SHOCK IMPEDANCES WERE NOTED BY THE PATIENT'S CLINIC. IT WAS NOTED THAT THE CLINIC WAS ALREADY AWARE OF THIS ONGOING ISSUE AND WAS ELECTING TO CONTINUE MONITORING THE PATIENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Additional Manufacturer Narrative
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Remedial action
- Other