inforMED
MalfunctionLRY

40 DEG UP-BITING LAMINECTOMY PUNCH 4MM WIDTH/330MM LENGTH

Received May 23, 2023 · Event occurred Apr 4, 2023

Report 8030965-2023-06496 · MDR key 16986048

Device

Generic name

Punch, Surgical

Manufacturer

Synthes Gmbh

Model number

389.41

Catalog number

389.41

Lot number

T165804

Product problems

  • Use of Device Problem
  • Compatibility Problem
  • Material Integrity Problem
  • Use of Device Problem
  • Compatibility Problem
  • Material Integrity Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE 40 DEG UP-BIT LAMNECT PNCH 4MM W/330MM L HAD THE UPPER SPRING BROKEN IN TWO PIECES, ONE FRAGMENT REMAINS ATTACHED TO THE HANDLE AND THE OTHER FRAGMENT WAS NOT RETURNED FOR EVALUATION. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. ENSURING PROPER HANDLING OF THE DEVICE IS RECOMMENDED TO AVOID BREAKAGE IN-SITU. A DIMENSIONAL INSPECTION FOR THE 40 DEG UP-BIT LAMNECT PNCH 4MM W/330MM L WAS NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION. A FUNCTIONAL TEST WAS UNABLE TO BE PERFORMED DUE TO THE BREAKAGE OF THE SPRING, HOWEVER, THIS CONDITION COULD HAVE CONTRIBUTED TO THE FUNCTIONAL ISSUES REPORTED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE 40 DEG UP-BIT LAMNECT PNCH 4MM W/330MM L WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW (DHR): PART NUMBER: 389.41. LOT NUMBER: T165804. MANUFACTURING SITE: TUTTLINGEN. RELEASE TO WAREHOUSE DATE: 18-APR-2018. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. THE RAW MATERIAL CERTIFICATE WAS REVIEWED AND THE USED MATERIAL WAS ACCORDING TO THE SPECIFICATION OF THE DEVICE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H6: MEDICAL DEVICE PROBLEM CODES A17 (COMPATIBILITY PROBLEM) AND A23 (USE OF DEVICE PROBLEM) ADDED.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. ADDITIONAL NARRATIVE: PART NUMBER: 389.41 LOT NUMBER: T165804 MANUFACTURING SITE: TUTTLINGEN RELEASE TO WAREHOUSE DATE: 18-APR-2018 THE PHOTO WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE FROM THE PHOTO. VISUAL ANALYSIS OF THE PHOTO REVEALED THAT 40 DEG UP-BIT LAMNECT PNCH 4MM W/330MM L HAD BROKEN ONE LEVER SPRING MATERIAL THAT ALLOWS THE HANDLE TO ARTICULATE. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED FOR 40 DEG UP-BIT LAMNECT PNCH 4MM W/330MM L. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. REPORTER IS A DISTRIBUTOR. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT ON APR 4, 2023, THE DISTRIBUTOR FOR THE SURGEON LISTED RECEIVED AN INSTRUMENT SET AND THE 4MM RONGEUR WAS BROKEN AT THE HANDLE BECAUSE IT WAS MISSING THE LEVER PIECE THAT ALLOWS THE HANDLES TO ARTICULATE. THIS DID NOT ALLOW THE INSTRUMENT TO FUNCTION AS INTENDED, SO IT WAS NOT USED DURING THE CASE. OTHER INSTRUMENTS WERE USED TO SUCCESSFULLY COMPLETE THE CASE. NO PATIENT INJURY OR HARM WAS INDICATED. THE SPREADER IN THEIR POSSESSION WAS LATER IDENTIFIED TO BE BROKEN AS WELL BECAUSE IT WOULD NOT RATCHET AS INTENDED DUE TO SPRING DAMAGE IN THE HANDLE COMPONENT. THIS REPORT IS FOR ONE (1) 40 DEG UP-BITING LAMINECTOMY PUNCH 4MM WIDTH/330MM LENGTH. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).